Table 3 Reliability and feasibility

From: Accuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: a prospective clinical trial

 

Per sensor

Initialization time

27 min

Total time until first displayed value

37.5 min [36/42]

Expected monitoring time after initialization

72 h

Actual monitoring time after initialization

37.9 h [23/71.3]

 Real-time data display

32.5 h [16/62.4]

 Percentage of real-time data display/expected monitoring time after initialization

45.1 %

 Percentage of real-time data display/actual monitoring time after initialization

84.8 % [67.9/93.8]

Data gaps after initialization

5 h [1.9/8.3]

 Percentage of data gaps/actual monitoring time after initialization

15.2 % [6.2/32.1]

 Number of performed calibrations

9.5 [6/13]

Reasons for the 11,568 min (192.8 h) of data gaps

Percentage of data gaps 11,568 min (192.8 h)

Percentage of monitoring time 80,223 min (1337.1 h)

1 Poor sensor signal (%)

23.3

3.4

2 Sensor failure (%)

15.0

2.2

3 Processor line error (%)

10.9

1.6

4 Disconnection (%)

15.6

2.3

5 Pending after reconnection (%)

3.4

0.5

6 Calibration required (%)

27.8

4.0

7 Others (%)

4.0

0.6

Device-related reasons (1–3) (%)

49.2

7.2

Not device related (4–7) (%)

50.8

7.4

Reasons for 21 prematurely sensor removals

Number of sensors

Percentage of all sensors (%)

Percentage of removed sensors (%)

Sensors 72 h completed

10

32.3

Sensors removed prematurely

21

67.7

100

1 Accidentally

7

22.6

33.3

2 Poor sensor signal during measurement

7

22.6

33.3

3 Poor sensor signal immediately after initialization

1

3.2

4.8

4 MRI

1

3.2

4.8

5 Discharge

1

3.2

4.8

6 Death

2

6.5

9.5

7 Others

2

6.5

9.5

Device-related reasons (1–3)

15

48.4

71.4

Not device related (4–7)

6

19.4

28.6

  1. n = 20 patients, 31 sensors
  2. Results are expressed as median with interquartile range or as absolute numbers with percentages
  3. MRI magnetic resonance imaging