Skip to main content

Table 2 Metrics of glucose control

From: Point and trend accuracy of a continuous intravenous microdialysis-based glucose-monitoring device in critically ill patients: a prospective study

Total number of measurements

594

Mean blood glucose level per patient, mg/dL, median [IQR; total range]

133 [118–140; 112–162]

Standard deviation of blood glucose level per patient, mg/dL, median [IQR; total range]

15 [11–18; 1–49]

Number of measurements per patient, median, [IQR; total range]

65 [26–80; 8–97]

Mild hyperglycemia 150–179 mg/dL in measurements, number (%)

62 (10)

Mild hyperglycemia 150–179 mg/dL in patients, number (%)

10 (91)

Severe hyperglycemia >180 mg/dL in measurements, number (%)

29 (5)

Severe hyperglycemia >180 mg/dL in patients, number (%)

3 (27)

Severe hypoglycemia ≤40 mg/dL in measurements, number (%)

0

Severe hypoglycemia ≤40 mg/dL in patients, number (%)

0

Mild hypoglycemia 41–70 mg/dL in measurements, number (%)

0

Mild hypoglycemia 41–70 mg/dL in patients, number (%)

0

  1. IQR interquartile range