Table 3 Reliability and safety of the CGM device

From: Point and trend accuracy of a continuous intravenous microdialysis-based glucose-monitoring device in critically ill patients: a prospective study

 

In total

Total number of sensors used

16

 

Per patient

Number of sensors used, median [IQR; total range]

1.0 [1.0–2.0; 1–2]

Total connection time, median [IQR; total range]

50.8 [13.5–54.7; 2.5–55.8] h

Real-time data, median [IQR; total range] (%)

50.8 [13.5–54.7; 2.5–55.8] h (100 %)

Time of skips in data acquisition

0.0 h

Percentage of time skips in data acquisition

0.0 h

Initialization time, median [IQR; total range]

42 [42–43; 40.8–62.4] min

Total start-up time, median [IQR; total range]

58 [56–67; 48–112.8] min

Number of calibrations needed before start, median [IQR; total range]

1.0 [1.0–1.3; 1–3]

Number of calibrations during duration of measurement, median [IQR; total range]

6.5 [2.0–7.0; 1–8]

Number of failed calibrations during duration of measurement, median [IQR; total range]

0.5 [0.0–2.0; 0–3]

  1. IQR, interquartile range; initialization time, total time from connecting the device, to being ready for calibration; total start-up time, total time from connecting the device, to displaying the first glucose value