From: A new global and comprehensive model for ICU ventilator performances evaluation
Device | Power on (%) | Start ventilation (%) | Inspiratory flow setting (%) | Ventilatory mode modification (%) | Cycling setting (%) | NIV mode activation (%) |
---|---|---|---|---|---|---|
Carefusion Avea | 92 | 100 | 100 | 100 | 75 | 75 |
Dräger V500 | 100 | 92 | 100 | 100 | 92 | 50 |
Covidien PB980 | 100 | 100 | 100 | 91 | 100 | 100 |
Philips V680 | 100 | 100 | 50§ | 91 | 91 | 40† |
Hamilton S1 | 91 | 91 | 18§ | 82 | 91 | 91 |
GE R860 | 100 | 100 | 100 | 91 | 100 | 91 |
Maquet Servo-U | 36* | 100 | 0¤ | 100 | 45‡ | 27† |
Device | Ventilator offset (%) | Alarm shut down (%) | Mode recognition (%) | Humidification system recognition (%) | Settings reading (%) | Overall |
---|---|---|---|---|---|---|
Carefusion Avea | 100 | 100 | 100 | 75 | 42 | 87%/85% |
Dräger V500 | 100 | 75 | 100 | 42 | 25 | 80%/75% |
Covidien PB980 | 73* | 64 | 100 | 73 | 45 | 86%/86% |
Philips V680 | 100 | 82 | 100 | 55 | 18 | 75%/70% |
Hamilton S1 | 100 | 91 | 100 | 36 | 27 | 74%/70% |
GE R860 | 100 | 64 | 100 | 9‖ | 27 | 80%/76% |
Maquet Servo-U | 100 | 82 | 100 | 0‖ | 45 | 58%#/55%‖ |