Table 1 Objective tasks completion rate

From: A new global and comprehensive model for ICU ventilator performances evaluation

Device

Power on (%)

Start ventilation (%)

Inspiratory flow setting (%)

Ventilatory mode modification (%)

Cycling setting (%)

NIV mode activation (%)

Carefusion Avea

92

100

100

100

75

75

Dräger V500

100

92

100

100

92

50

Covidien PB980

100

100

100

91

100

100

Philips V680

100

100

50§

91

91

40

Hamilton S1

91

91

18§

82

91

91

GE R860

100

100

100

91

100

91

Maquet Servo-U

36*

100

0¤

100

45

27

Device

Ventilator offset (%)

Alarm shut down (%)

Mode recognition (%)

Humidification system recognition (%)

Settings reading (%)

Overall

Carefusion Avea

100

100

100

75

42

87%/85%

Dräger V500

100

75

100

42

25

80%/75%

Covidien PB980

73*

64

100

73

45

86%/86%

Philips V680

100

82

100

55

18

75%/70%

Hamilton S1

100

91

100

36

27

74%/70%

GE R860

100

64

100

9

27

80%/76%

Maquet Servo-U

100

82

100

0

45

58%#/55%

  1. Results are presented as the different objective tasks success rate, expressed as the percentage of successful attempts. ICU’s physicians with mechanical ventilation’s knowledge had to complete 11 specific tasks of variable clinical importance for each ventilator, 4 mainly dedicated to monitoring and 7 to setting. Overall provides results of the entire bunch of test (first value), or after powering on/switching off tasks exclusion (second value)
  2. Among all ventilators, our reference device the Avea had the better success rate. In between our comparison of the six ventilators, Covidien PB980 had the better results and the Servo-U the worst. A minority of users could power on Servo-U and always over the predefined 120 s time range. Settings reading represented the more difficult task, whatever the device, while their terminology was highly different from one device to the other
  3. p < 0.01 as compared to others. § p < 0.001 as compared to Avea, V500 and PB980. ¤ p < 0.01 as compared to V680.  p < 0.05 as compared to others except Avea.  p < 0.05 as compared to PB980 S1 and R860.  p < 0.001 as compared to Avea and PB980. # p < 0.01 as compared to others, except S1 and V680