Fig. 2 | Annals of Intensive Care

Fig. 2

From: Validation of transpulmonary thermodilution variables in hemodynamically stable patients with heart diseases

Fig. 2

Good correlation between cardiac output measurements by transpulmonary thermodilution and pulmonary artery catheter was observed [r = 0.69, p < 0.0001, bias 0.9 l min−1 (extended black line in the Bland-Altman plot), precision 0.8 l min−1, level of agreement −0.7 to 2.6 l min−1 (dashed black line)]. The shaded area in the linear correlation plot denotes the fitted linear model’s 95% confidence interval. AS aortic valve stenosis, DCM dilated cardiomyopathy