Table 2 Mean volume errors and Positive End Expiratory Pressure 1 (PEEP) measured for each ventilator

From: Reliability and limits of transport-ventilators to safely ventilate severe patients in special surge situations

  Engström Carestation Osiris 3 Oxylog 3000 Monnal T60 Elisee 350
Vteerror global [%] 2.9 ± 2.2 3.6 ± 3.9 2.5 ± 2.1 5.4 ± 2.7 (*) 8.8 ± 4.8 (*)
Vteerror 100% FiO2 [%] 1.0 ± 0.7 3.7 ± 3.7 (*) 2.0 ± 1.2 (*) 3.3 ± 1.4 (*) 11.9 ± 4.1 (*)
Vteerror 70% FiO2 [%] 4.9 ± 1.3 3.5 ± 4.2 2.9 ± 2.7 (*) 7.5 ± 2.0 (*) 5.9 ± 3.5
Vteerror C50 [%] 3.3 ± 2.7 3.4 ± 4.0 1.6 ± 1.1 (*) 5.1 ± 3.1 (*) 10.4 ± 5.3 (*)
Vteerror C20 [%] 2.6 ± 1.6 3.8 ± 3.8 3.4 ± 2.6 5.7 ± 2.2 (*) 7.0 ± 3.7 (*)
Mean PEEP 10 [cmH2O] 9.9 ± 0.2 10.6 ± 0.6 (*) 11.5 ± 0.3 (*) 9.5 ± 0.5 (*) 10.3 ± 0.1 (*)
Mean PEEP 15 [cmH2O] 15.1 ± 0.2 14.9 ± 0.6 (*) 15.4 ± 2.2 14.7 ± 0.1 (*) 15.4 ± 0.2 (*)
  1. Vteerror global = mean volume error including all conditions of resistance (15 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for both 100% FiO2 and 70% FiO2. Vteerror 100% FiO2= mean volume error including all conditions of resistance (15 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for 100% FiO2, Vteerror 70% FiO2= mean volume error including all conditions of resistance (15–12 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for 70% FiO2. Vteerror C50 = mean volume error including all conditions of 1 resistance (15 cmH2O/l/s), FiO2 (100–70%) and PEEP (10–15 cmH2O) for a compliance of 50 ml/cmH2O. teerror C20 = mean volume error including all conditions of resistance (15 cmH2O/l/s), FiO2 (100–70%) and PEEP (10–15 cmH2O) for a compliance of 20 ml/cmH2O, Mean PEEP 10 = mean PEEP measured when PEEP was set at 10 cmH2O including all conditions of resistance (15 cmH2O/l/s) and compliance (20–50 ml/cmH2O) for both 100% FiO2 and 70% FiO2. Mean PEEP 15 = mean PEEP measured when PEEP was set at 15 cmH2O including all conditions of resistance (15 cmH2O/l/s) and compliance (20–50 ml/cmH2O) for both 100% FiO2 and 70% FiO2
  2. *p < 0.005 when comparing each transport ventilator with the Engstrom ICU ventilator (ANOVA test: global F was significant)
\