Table 2 Mean volume errors and Positive End Expiratory Pressure 1 (PEEP) measured for each ventilator

From: Reliability and limits of transport-ventilators to safely ventilate severe patients in special surge situations

 

Engström Carestation

Osiris 3

Oxylog 3000

Monnal T60

Elisee 350

Vteerror global [%]

2.9 ± 2.2

3.6 ± 3.9

2.5 ± 2.1

5.4 ± 2.7 (*)

8.8 ± 4.8 (*)

Vteerror 100% FiO2 [%]

1.0 ± 0.7

3.7 ± 3.7 (*)

2.0 ± 1.2 (*)

3.3 ± 1.4 (*)

11.9 ± 4.1 (*)

Vteerror 70% FiO2 [%]

4.9 ± 1.3

3.5 ± 4.2

2.9 ± 2.7 (*)

7.5 ± 2.0 (*)

5.9 ± 3.5

Vteerror C50 [%]

3.3 ± 2.7

3.4 ± 4.0

1.6 ± 1.1 (*)

5.1 ± 3.1 (*)

10.4 ± 5.3 (*)

Vteerror C20 [%]

2.6 ± 1.6

3.8 ± 3.8

3.4 ± 2.6

5.7 ± 2.2 (*)

7.0 ± 3.7 (*)

Mean PEEP 10 [cmH2O]

9.9 ± 0.2

10.6 ± 0.6 (*)

11.5 ± 0.3 (*)

9.5 ± 0.5 (*)

10.3 ± 0.1 (*)

Mean PEEP 15 [cmH2O]

15.1 ± 0.2

14.9 ± 0.6 (*)

15.4 ± 2.2

14.7 ± 0.1 (*)

15.4 ± 0.2 (*)

  1. Vteerror global = mean volume error including all conditions of resistance (15 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for both 100% FiO2 and 70% FiO2. Vteerror 100% FiO2= mean volume error including all conditions of resistance (15 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for 100% FiO2, Vteerror 70% FiO2= mean volume error including all conditions of resistance (15–12 cmH2O/l/s), compliance (20–50 ml/cmH2O) and PEEP (10–15 cmH2O) for 70% FiO2. Vteerror C50 = mean volume error including all conditions of 1 resistance (15 cmH2O/l/s), FiO2 (100–70%) and PEEP (10–15 cmH2O) for a compliance of 50 ml/cmH2O. teerror C20 = mean volume error including all conditions of resistance (15 cmH2O/l/s), FiO2 (100–70%) and PEEP (10–15 cmH2O) for a compliance of 20 ml/cmH2O, Mean PEEP 10 = mean PEEP measured when PEEP was set at 10 cmH2O including all conditions of resistance (15 cmH2O/l/s) and compliance (20–50 ml/cmH2O) for both 100% FiO2 and 70% FiO2. Mean PEEP 15 = mean PEEP measured when PEEP was set at 15 cmH2O including all conditions of resistance (15 cmH2O/l/s) and compliance (20–50 ml/cmH2O) for both 100% FiO2 and 70% FiO2
  2. *p < 0.005 when comparing each transport ventilator with the Engstrom ICU ventilator (ANOVA test: global F was significant)