Table 5 Automated injection phase: Comparison between triple-lumen 6Fr PICC vs. CICC

From: Are single-lumen 5Fr and triple-lumen 6Fr PICCs suitable for hemodynamic assessment by trans-pulmonary thermodilution? A pilot study

Variables

6 Fr triple-lumen PICC

CICC

p value

Bias [LoA]

Bias,  %

Error,  %

Measurements, n

40

40

CI, L/min/m2

3.4 (0.7)

3.0 (0.7)

0.012

0.39 [− 0.05; 0.83]

12.5

13.8

GEDVI, mL/m2

689 (117)

622 (89)

0.006

66.5 [− 40; 173]

9.7

16.2

EVLWI, mL/kg

14.3 (5.0)

12.7 (4.5)

0.134

1.6 [− 0.4; 3.7]

12

15.6

SVI, mL/m2

44.6 (8.4)

39.4 (7.5)

0.004

5.2 [− 1.3; 11.7]

12.5

15.5

CVP, mmHg

12 (5.3)

11.1 (4.9)

0.458

ΔT,  °C

0.30 (0.07)

0.34 (0.08)

0.008

− 0.05 [− 0.09; 0.00]

15.6

Psi

120

115

< 0.0001

  1. Data are reported as mean (standard deviation)
  2. CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; Psi: pound per square inch; SVI: stroke volume index; ΔT: delta temperature