Fig. 4 | Annals of Intensive Care

Fig. 4

From: Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU

Fig. 4

Effect of inhaled sedation devices on end-tidal CO2. The effects of inhaled sedation devices on end-tidal CO2 were first analyzed under ambient temperature pressure dry (ATPD) conditions (bench model #1) and then under body temperature pression saturated (BTPS) conditions (humid conditions, bench model #2). Experiments in ATPD and BTPS conditions were independent Dashed redlines indicate the baseline value of end-tidal CO2 (40 mmHg). Error bars indicate standard deviation. CTRL: Control group (green circle, no device); MIRUS: Mirus™ device (blue square); ANA-100: AnaConDa-100 mL device (orange triangle); ANA-50: AnaConDa-50 mL device (yellow triangle). *p < 0.05 vs CTRL, †p < 0.05 vs MIRUS, ‡p < 0.05 vs ANA-100, §p < 0.05 vs ANA-50