Table 1 Effect of inhaled sedation devices on ventilatory pressures, driving pressure, and Costa index at lung compliances of 60, 40 and 20 mL/cmH2O

From: Effects on mechanical power of different devices used for inhaled sedation in a bench model of protective ventilation in ICU

 

PEEP 5 cmH2O

PEEP 15 cm H2O

Compliance 60 mL/cm H2O

Baseline

CTRL

MIRUS

ANA-100

ANA-50

CTRL

MIRUS

ANA-100

ANA-50

Ppeak

16

16

16

16

26

26a

27a

27a

Pplat

11

12a

12a

12a

21

22a

23a

23a

PEEPt

5

6a

6a

6a

15

16a

16a

17a

ΔP

6

6a

6a

6a

6

6a

6a

6a

4×ΔP + RR

44

45ad

45a

45ab

44

45ad

45a

45ab

VT correction

 Ppeak

16

18a

19a

19a

26

29a

30a

29

 Pplat

11

15a

16a

15a

21

25a

26a

25a

 PEEPt

5

6a

6a

6a

15

16a

17a

17a

 ΔP

6

9a

9a

8a

6

8a

9a

8a

 Final VT

400

560

580

520

400

580

580

520

 4 × ΔP + RR

44

55acd

57abd

53abc

44

54acd

57abd

53abc

RR correction

 Ppeak

16

19a

20a

19a

26

30a

30a

28a

 Pplat

11

15a

16a

14a

21

26a

26a

24a

 PEEPt

5

8a

8a

7a

15

19a

19a

17a

 ΔP

6

7a

8a

7a

6

7a

7a

7a

 Final RR

20

32

35

29

20

35

35

26

 4 × ΔP + RR

44

61a

67a

57a

44

65a

67a

57a

 

Compliance 40 mL/cmH2O

Baseline

CTRL

MIRUS

ANA-100

ANA-50

CTRL

MIRUS

ANA-100

ANA-50

Ppeak

18

18

19

19

29

29

29

29

Pplat

14

15a

15a

15a

24

25a

25a

25a

PEEPt

5

6a

6a

6a

15

16a

16a

16a

ΔP

9

9a

9a

9a

9

9a

9a

9a

4 × ΔP + RR

55

56acd

56abd

57abc

55

56acd

56abd

57abc

VT correction

 Ppeak

18

22a

23a

22a

29

32a

33a

32a

 Pplat

14

19a

20a

18a

24

28a

30a

28a

 PEEPt

5

6a

6a

6a

15

16a

17a

16a

 ΔP

9

13a

14a

12a

9

12a

13a

12a

 Final VT

400

580

580

520

400

580

580

520

 4xΔP + RR

55

71acd

74abd

68abc

55

71acd

74abd

68abc

RR correction

 Ppeak

18

21a

21a

20a

29

32a

32a

31a

 Pplat

14

18a

17a

17a

24

28a

28a

27a

 PEEPt

5

8a

7a

7a

15

18a

18a

17a

 ΔP

9

10a

10a

10a

9

10a

10a

10a

 Final RR

20

35

35

29

20

35

35

29

 4xΔP + RR

55

75acd

76abd

68abc

55

75acd

76abd

68abc

 

Compliance 20 mL/cm H2O

Baseline

CTRL

MIRUS

ANA-100

ANA-50

CTRL

MIRUS

ANA-100

ANA-50

 Ppeak

26

25

26

26

36

36

37a

37

 Pplat

22

23a

23a

23a

32

33a

34a

34a

 PEEPt

5

5a

6a

6a

15

16a

16a

16a

 ΔP

17

17a

17a

17a

17

17a

17a

17a

 4 × ΔP + RR

88

88cd

90ab

90ab

88

88cd

90ab

90ab

VT correction

 Ppeak

26

31a

34a

32a

36

41a

43a

41a

 Pplat

22

28a

32a

29a

32

38a

41a

38a

 PEEPt

5

5a

6a

6a

15

16a

16a

16a

 ΔP

17

22a

25a

23a

17

22a

24a

22a

 Final VT

400

580

580

520

400

580

580

520

 4xΔP + RR

88

109acd

123abd

114abc

88

109acd

123abd

114abc

RR correction

 Ppeak

26

27a

27a

27a

36

37a

38a

37a

 Pplat

22

24a

24a

24a

32

34a

35a

34a

 PEEPt

5

6a

6a

6a

15

17a

17a

16a

 ΔP

17

17a

18a

18a

17

18a

18a

18a

 Final RR

20

35

35

29

20

35

35

26

 4 × ΔP + RR

88

106acd

107abd

102abc

88

106acd

107abd

102abc

  1. Data are expressed as mean and standard deviations are not shown because they are equal to 0 for each parameter
  2. CTRL Control group (no device, values reported in bold), MIRUS Mirus device, ANA-100 AnaConDa device with 100 mL volumetric dead space; ANA-50: AnaConDa device with 50 mL volumetric dead space; PPeak: peak pressure (cmH2O); PPlat: plateau pressure (cmH2O); PEEPt: total positive end-expiratory pressure (cmH2O); ΔP: driving pressure (cmH2O); 4×ΔP + RR: Costa index (arbitrary units)
  3. Final VT: tidal volume (expressed in mL) set on the ventilator to maintain end-tidal CO2 at 40 ± 2 mmHg; Final RR: respiratory rate (breaths/minute) set on the ventilator to maintain end-tidal CO2 at 40 ± 2 mmHg
  4. ap < 0.01 vs. CTRL
  5. bp < 0.01 vs. MIRUS
  6. cp < 0.01 vs. ANA-100
  7. dp < 0.01 vs. ANA-50