Skip to main content

Table 2 Factors potentially influencing endotracheal intubation time

From: Usefulness of video-laryngoscopy with the Airway Scope for intubation performance and learning: an experimental manikin controlled study

 

All attempts (n = 1200)

Univariable analysis*

Multivariable analysis**

ETI Time (s)

95 % CI limits

p value

ETI time at first attempt for novices 1 (s)

Time multiplier***

95 % CI limits

p value

Number of attempts, n

Sample arithmetic mean time (s)

Lower

Upper

Lower

Upper

ETI device

1200

19.63

         

 DLM

600

24.52

20.44

16.18

25.83

<0.0001

35.65

29.61

42.93

<0.0001

 AWS

600

14.73

12.72

10.08

16.06

 

22.51

18.75

27.02

 

Operator sub-group (device order)

 Experts (DLM then AWS)

400

12.53

11.01

9.28

13.07

<0.0001

0.54

0.49

0.58

<0.0001

 Novices 1 (DLM then AWS)

400

25.35

20.51

17.29

24.32

 

1.00

 Novices 2 (AWS then DLM)

400

21.00

18.37

15.49

21.79

0.02

0.83

0.77

0.90

<0.0001

Attempt number in the series of 10 attempts

 1

120

24.92

20.99

16.42

26.84

Reference

1.00

 2

120

23.63

19.44

15.16

24.92

0.16

0.91

0.83

1.01

0.08

 3

120

20.62

17.09

13.33

21.93

0.003

0.80

0.71

0.89

0.0001

 4

120

19.56

15.92

12.41

20.42

<0.0001

0.74

0.66

0.84

<0.0001

 5

120

18.76

15.22

11.87

19.51

<0.0001

0.71

0.63

0.80

<0.0001

 6

120

18.22

14.87

11.59

19.09

<0.0001

0.69

0.62

0.77

<0.0001

 7

120

18.65

15.34

11.94

19.72

<0.0001

0.71

0.63

0.80

<0.0001

 8

120

17.81

14.59

11.35

18.75

<0.0001

0.68

0.60

0.76

<0.0001

 9

120

17.03

14.06

10.96

18.03

<0.0001

0.65

0.58

0.73

<0.0001

 10

120

17.06

13.69

10.61

17.66

<0.0001

0.63

0.57

0.70

<0.0001

  1. ETI endotracheal intubation, DLM direct laryngoscopy with Macintosh, AWS video-laryngoscopy with Airway Scope®
  2. * Univariable analysis, all the 1200 attempts have been analyzed separately according to 3 variables: ETI device (DLM vs. AWS), operator sub-group (experts vs. novices 1 and novices 1 vs. novices 2) and the attempt number in the series of 10 attempts
  3. ** Novices 1 at the first attempt serve as reference situation of the multivariable analysis, i.e., the simultaneous analysis of 3 variables: ETI device (DLM vs. AWS), operator sub-group (experts vs. novices 1 and novices 1 vs. novices 2) and the attempt number in the series of 10 attempts
  4. *** Multiplicative factor allowing to estimate ETI time from the novices 1 ETI time for the others situations. For example, ETI time for an expert using AWS at first attempt is estimated to be 12.16 s, i.e., 22.51 × 0.54 × 1