Fig. 1 | Annals of Intensive Care

Fig. 1

From: Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device

Fig. 1

Correlation and Bland–Altman plot for reference and estimated respiratory rate during the retrospective evaluation phase. Reference respiratory rate (RR) values are provided as the chronometric measurement measured over 30 s under spontaneous ventilation or as the ventilatory rate for patients under ventilatory assistance. Estimated RR-PPG was performed using the plethysmographic curve analysis. On the 201 ICU patients’ recordings, retrospective estimation of the RR-PPG was highly correlated with the reference method (r = 0.95; p < 0.0001). Comparison of the methods using the Bland–Altman method depicted a low bias (0.1 c/min) and deviation (< ± 3.5 c/min)