Table 3 Manual phase: Comparison between triple-lumen 6Fr PICC vs. CICC

From: Are single-lumen 5Fr and triple-lumen 6Fr PICCs suitable for hemodynamic assessment by trans-pulmonary thermodilution? A pilot study

Variables

6 Fr triple-lumen PICC

CICC

p value

Bias [LoA]

Bias,  %

Error,  %

Measurements, n

40

40

CI, L/min/m2

3.3 (0.8)

3.0 (0.7)

0.107

0.28 [− 0.32; 0.88]

8.5

19

GEDVI, mL/m2

685 (133)

632 (102)

0.05

52.8 [− 75.7; 181]

7.4

19.5

EVLWI, mL/kg

14.0 (5.1)

12.2 (4.9)

0.178

1.5 [− 2.2; 5.2]

12.8

28.2

SVI, mL/m2

44.4 (10.4)

40.6 (8.5)

0.077

3.8 [− 4.2; 1.8]

8.5

18.8

CVP, mmHg

11.3 (4.8)

11.7 (5.5)

0.764

ΔT,  °C

0.33 (0.07)

0.34 (0.08)

0.514

− 0.01 [− 0.06; 0.03]

12.9

  1. Data are reported as mean (standard deviation)
  2. CI: cardiac index; CICC: centrally inserted central catheter; CVP: central venous pressure; EVLWI: extra-vascular lung water index; GEDVI: global end-diastolic volume index; LoA: limits of agreement; PICC: peripherally inserted central catheter; SVI: stroke volume index; ΔT: delta temperature