Sleep quality in mechanically ventilated patients: comparison between NAVA and PSV modes
Annals of Intensive Care volume 1, Article number: 42 (2011)
Abstract
Background
Mechanical ventilation seems to occupy a major source in alteration in the quality and quantity of sleep among patients in intensive care. Quality of sleep is negatively affected with frequent patient-ventilator asynchronies and more specifically with modes of ventilation. The quality of sleep among ventilated patients seems to be related in part to the alteration between the capacities of the ventilator to meet patient demand. The objective of this study was to compare the impact of two modes of ventilation and patient-ventilator interaction on sleep architecture.
Methods
Prospective, comparative crossover study in 14 conscious, nonsedated, mechanically ventilated adults, during weaning in a university hospital medical intensive care unit. Patients were successively ventilated in a random ordered cross-over sequence with neurally adjusted ventilatory assist (NAVA) and pressure support ventilation (PSV). Sleep polysomnography was performed during four 4-hour periods, two with each mode in random order.
Results
The tracings of the flow, airway pressure, and electrical activity of the diaphragm were used to diagnose central apneas and ineffective efforts. The main abnormalities were a low percentage of rapid eye movement (REM) sleep, for a median (25th-75th percentiles) of 11.5% (range, 8-20%) of total sleep, and a highly fragmented sleep with 25 arousals and awakenings per hour of sleep. Proportions of REM sleep duration were different in the two ventilatory modes (4.5% (range, 3-11%) in PSV and 16.5% (range, 13-29%) during NAVA (p = 0.001)), as well as the fragmentation index, with 40 ± 20 arousals and awakenings per hour in PSV and 16 ± 9 during NAVA (p = 0.001). There were large differences in ineffective efforts (24 ± 23 per hour of sleep in PSV, and 0 during NAVA) and episodes of central apnea (10.5 ± 11 in PSV vs. 0 during NAVA). Minute ventilation was similar in both modes.
Conclusions
NAVA improves the quality of sleep over PSV in terms of REM sleep, fragmentation index, and ineffective efforts in a nonsedated adult population.
Background
S leep is severely disturbed in mechanically ventilated ICU patients [1–3]. Sleep alterations are known to have deleterious consequences in healthy subjects, but the paucity of data in the literature [4–7] makes it difficult to determine the impact of sleep abnormalities in ICU patients. Intensive care unit (ICU) patients present disrupted sleep with reduced sleep efficiency and a decrease in slow wave sleep and rapid eye movement (REM) sleep [8–10]. Furthermore, polysomnographic studies performed on mechanically ventilated ICU patients have demonstrated an increase in sleep fragmentation, a reduction in slow-wave and REM sleep, and an abnormal distribution of sleep, because almost half of the total sleep time occurred during the daytime [11–13]. In the Freedman et al. study [14], noise was considered a nuisance for the patients questioned; the most annoying noises were alarms and caregivers' conversations. When the same authors simultaneously recorded noise and microarousal, they identified an association between arousal and noise in only 11-17% of the cases [11]. This percentage is confirmed by Gabor et al. [3] where 21% of the arousal interruptions were explained by loud noises and 7% to patients' care. Seventy-eight percent of the microarousals were not associated with environment noises, suggesting other causes, such as patient/ventilator asynchrony [3, 14].
The effects of assist control ventilation (ACV) and pressure support ventilation (PSV) on sleep fragmentation have been examined in critically ill patients receiving mechanical ventilation [15], where PSV mode was associated with increases in the number of central apneas and subsequent sleep fragmentation compared with AVC. Furthermore, the study suggested that PSV by itself or an excess of ventilator assistance with PSV could have caused such sleep alterations. Indeed, ventilatory settings adjusted during wakefulness may become excessive during sleep, as the patients' ventilatory demand is reduced while asleep [16]. Whether these results can be explained by the ventilatory mode itself or how it was adjusted is an important issue, because hyperventilation and patient ventilator asynchrony may result from PSV as well as ACV in mechanically ventilated ICU patients [17]. Fanfulla et al. [18] compared two ventilatory settings in nine patients under long-term PSV for neuromuscular disease. The initial setting was set according to clinical parameters, and the second setting was adjusted with measurement of esophageal pressure (physiological setting) to optimize patient effort. The physiological setting improved the duration and quality of sleep, decreased episodes of apnea, and the amount of inefficient efforts for ventilator triggering [18]. The level of pressure support and PEEP tended to decrease, with a lowering of intrinsic PEEP and patient-ventilator asynchronies. A recent study by Cabello et al. [19] compared the impact of three modes of ventilation (AVC, PSV, and SmartCare™) on the quality of sleep in alert and nonsedated patients, and no difference for the architecture, fragmentation, and duration of sleep was found among the three modes.
Our hypothesis is that NAVA ventilation is superior to PSV by allowing optimal patient-ventilator synchrony and thereby decreasing sleep fragmentation.
Methods
This study was approved by the Ethics Committee of the Hôpital du Sacré-Coeur de Montréal, and patients or their surrogates gave written informed consent.
Patients
This physiologic study was conducted in a 22-bed medical ICU during a 12-month period. The weaning phase of mechanical ventilation was chosen because patient-ventilator asynchrony is common when patients are spontaneously triggering breaths. The inclusion criteria required that the patient was conscious, free from sedation and opiate analgesia for ≥ 24 hours, and ventilated in PSV mode with an FIO2 < 60%, PEEP = 5 cmH2O, and SpO2 ≥ 90%. Exclusion criteria consisted of the presence of a central nervous system disorder, Glasgow Coma Scale score < 11, hemodynamic instability, renal and/or hepatic insufficiency, and ongoing sepsis.
Methods
All patients were ventilated through an endotracheal tube or a tracheostomy; once they met the inclusion criteria, they were connected to a Servo i ventilator (Maquet critical Care, Sölna, Sweden), equipped with a neurally adjusted ventilator assist system (NAVA). The electrical activity of the diaphragm (EAdi) is captured with the EAdi catheter (Maquet Critical Care, Sölna, Sweden) consisting of a 16-Fr gastric tube equipped with electrodes. End-tidal CO2 was monitored with the Servo-i Volumetric CO2 module. The two different ventilatory modes were delivered in a randomized order using a closed-envelope technique during four periods of 4 hours: a daytime period from 7 to 11 a.m. and 12 to 4 p.m., and a nocturnal period from 10 p.m. to 2 a.m. and 3 to 7 a.m. To prevent possible data contamination from the previous mode of ventilation, a 1-hour washout period after a ventilator change was introduced before data acquisition (Figure 1; Study Protocol).
During periods of wakefulness, PSV and NAVA were clinically adjusted by the attending physician to obtain a tidal volume of 8 mL/kg of predicted body weight and a respiratory rate ≤ 35 breaths/min. For both modes of ventilation, inspiratory triggering sensitivity was set at thresholds that would not allow auto-triggering for both modes of ventilation: 0.5 mV in NAVA and 5 in PSV.
EEG was recorded from standard locations: left frontal/right mastoïd reference (F3/M2 or F3/A2), right frontal/left mastoid reference (F4/M1or F4/A1), left central/right mastoïd reference (C3/M2 or C3/A2), right central/left mastoïd reference (C4/M1 or C4/A1), left occipital/right mastoïd reference (O1/M2 or O1/A2), and right occipital/left mastoïd reference (O2/M1or O2/A1), according to the International 10-20 System for electrode placement [20]. The standard reference used was the left mastoid lead [20]. Two electro-oculogram and three chin electromyogram leads were used to score REM and non-REM sleep. The electroencephalogram, the right and left electro-oculogram, and the submental electromyogram signals were amplified and recorded in the data acquisition system (Alice 5 polysomnography system using Alice® Sleepware™ 2.5 software, Respironics, Nantes, France).
Sleep recordings were manually read and scored by an independent pulmonologist blinded to the study, using the criteria of Rechtschaffen and Kales [21, 22] and the criteria of the American Sleep Disorder Association for arousals and awakenings [23, 24]. Diagnosis of central apnea was based on international recommendations [24]. The diagnosis of central apnea is characterized by absent breathing and respiratory effort for a period of at least 10 seconds. Arousals and awakenings were considered secondary to apnea when occurring within three cycles and/or 15 sec after a respiratory event [25, 26]. Ineffective efforts were defined as an inspiratory effort observed by a peak electrical activity of the diaphragm (EAdi peak) without a simultaneously triggered ventilator cycle. Airflow, Paw, and EAdi were acquired from the ventilator through a RS232 interface at a sampling rate of 100 Hz, recorded by a dedicated software (Nava Tracker V. 2.0, Maquet Critical Care, Sölna, Sweden), and an analyzer using software Analysis V 1.0 (Maquet Critical Care) and a customized software based for Microsoft Excel. An arousal or awakening event was considered secondary to ineffective triggering when it occurred within 15 seconds after the asynchrony [19].
Noise was measured with a portable noise meter at the level of patient's head (Quest Technologies, Oconomowoc, WI). Arousals and awakenings were associated with the noise when they occurred 3 seconds after or within noise increase ≥10 dB [3, 11]. Inspiratory trigger delay was calculated as the time difference between the onset of EAdi peak and Paw inspiratory swings. Cycling-off delay was calculated as the time difference between the end of the inspiratory EAdi peak deflection and the onset of expiratory flow.
Statistics analysis
Statistical analysis was performed using SPSS statistical software (SPSS 17.0). Continuous variables were expressed as median (25th-75th percentile) or mean ± SD. Data were compared using the general linear model for repeated measures (GLM). The small sample of patients led us to use Wilcoxon's t test for paired samples, and the p values for multiple comparisons were corrected for the Bonferroni inequality. A two-tailed p value < 0.05, corrected as needed, was retained to indicate statistical significance.
Results
Patients
Fourteen patients were selected and none were excluded during the study. Their main characteristics are shown in Table 1. Acute respiratory failure was the most frequent reason to initiate mechanical ventilation in ten patients, postoperative complications in three patients, and septic shock in one patient.
Sleep recordings
All patients completed the study, and recordings were well tolerated. Individual sleep data are shown in Table 2. The median total sleep time was 564 (range, 391-722) minutes. The median sleep efficiency (i.e., the percentage of sleep during the study) was 59% (range, 41-75%). The main abnormalities observed on each patient were a diminished percentage of REM sleep, counting for only 11.5% (range, 8-20%) of total sleep time, and a high fragmentation index with 25 arousals and awakenings per hour (range, 18-51). Although interindividual variability was large, the median quantity of slow-wave sleep (stages 3 and 4 or NREM3 stage) was normal, with a median of 18.5 (range, 11.5-22; Table 2).
Ventilatory modes and sleep distribution
Sleep efficiency and architecture appeared very different for both modes of ventilation (NAVA and PSV). Stage 1 (NREM1 stage) lasted longer during PSV compared with NAVA 7.5% (range, 4-15%) vs. 4% (range, 3-5%; p = 0.006). Stage 2 (NREM2 stage) also lasted longer in PSV than NAVA 68% (range, 66-75%) vs. 55% (range, 52-58%; p = 0.001). Stage 3-4 (NREM3 stage) was shorter in PSV as opposed to NAVA 16.5% (range, 17-20%) vs. 20.5% (range, 16-25%; p = 0.001). REM stage (R stage) was much shorter in PSV than in NAVA 4.5% (range, 3-11%) vs. 16.5% (range, 13-29%; p = 0.001). The fragmentation index was different between the two ventilation modes, with 40 ± 20 arousals and awakenings per hour in PSV and 16 ± 9 during NAVA (p = 0.001; Figure 2 Sleep stage (percent of total sleep) during two ventilatory modes; Table 3).
Minute ventilation did not significantly differ between PSV and NAVA with median values of 9.8 L/min (range, 8.0-10.9), and 9.6 L/min (range, 7.5-11.0) respectively (p = 0.51). The median respiratory rates were 17 breaths/min (range, 14-21), and 20 breaths/min (range, 15-23) during PSV and NAVA (p = 0.14). Median tidal volume was 420 mL (8.1 mL/Kg of predicted body weight; range, 375-479 mL), and 378 mL (7.3 mL/Kg of predicted body weight; range, 370-448 mL) during PSV and NAVA, respectively (p = 0.36). The mean PSV level was 15 ± 5 cmH2O, and the mean NAVA level was 1.6 ± 1.4 cmH2O/μV. Positive end-expiratory pressure was kept at 5 cmH2O for all patients.
Apneas and ineffective efforts
Ten of the 14 patients presented sleep apnea, and 11 exhibited ineffective efforts. The mean index of sleep apneas (number of apneas per hour of sleep) was 10.5 ± 11 apneas during PSV and 0 during NAVA (p = 0.005) and ineffective efforts (number of ineffective efforts per hour of sleep) was 24 ± 23 ineffective efforts during PSV and 0 during NAVA (p = 0.001). Over-assistance during sleep is sensed on the previous three cycles preceding central apnea. Tidal volume and minute ventilation increased, whereas ETCO2 and EAdi decreased over the three cycles preceding central apnea Table 4.
Trigger delay and cycling-off delay
During N-REM sleep in PSV, the trigger delay increased on average by 80 ± 26 (msec) during stage 1 versus 158 ± 42 (msec) during stage 3 and 4. The expiratory trigger (cycling-off) increased in PSV by 158 ± 103 (msec) and 258 ± 87 (msec) during stage 1 and stages 3 and 4, respectively. In NAVA, the trigger delay remained stable during sleep, 68 ± 24 (msec) during stage 1 and 72 ± 32 (msec) during stages 3 and 4. The expiratory trigger also remained stable in NAVA: 39 ± 28 (msec) during stage 1 and 41 ± 34 (msec) during stages 3 and 4.
Noise
In ICU, we recorded the average baseline ambient noise level and evaluated arousals from this baseline to a peak noise level ≥ 10 dB above ambient noise level. The mean noise level was recorded at 64 ± 8 dB, with the peak level recorded at 111 dB and the minimal level at 52 dB. No differences were observed between the two different ventilatory modes concerning the index of fragmentation associated with noise: 7.5 ± 3 during PSV and 6 ± 3.5 during NAVA (p = 0.19). These data indicate that 18% during PSV and 21% during NAVA of the fragmentation was associated with sudden increases in noise.
Sleep distribution among study periods
The cross-over pattern was balanced with an equal number of patients from each sequence initiating the rotation. Independent of the ventilatory mode, sleep efficiency and sleep architecture had a significantly different distribution based on the study period considered (Figure 3--sleep stage (percent of total sleep) during the four daily time periods). Sleep efficiency was the same in the two daytime periods (2 periods during the day): 52% (range, 26-67%) during the first day period (7 h-11 h a.m.) and 51.5% (range, 27-67%) during the second day period (12 h-4 h p.m.; p = 0.18). Sleep efficiency also did not differ between the two night periods: 65.5% (range, 37-82%) during the first night period (10 h p.m.-2 h a.m.) and 65% (range, 45-82.5%) during the second nighttime period (3 h-7 h a.m.; p = 0.11).
There was no statistical difference between stage 1 and 2 recording periods. A greater duration of slow-wave sleep (stage 3-4) was found during the first nocturnal period with a median percentage 22.5% (range, 20-33.5%) vs. 15.5% (range, 7-19.5%) during first day period (p = 0.03), vs. 15% (range, 7-18%) during second day period (p = 0.01) and vs. 18% (range, 13-21%) during second nighttime period (p = 0.001).
The proportion of REM sleep was longer during the second nocturnal period, with a median percentage of 16.5% (range, 15-25%) vs. 11.5% (range, 5-15%) during first day period (p = 0.001) vs. 9% (range, 5-15%) during second day period (p = 0.001) and vs. 10.5% (range, 7-21%) during first nighttime period (p = 0.02). The fragmentation index did not differ with 26 (range, 20-65) arousals and awakenings/hour during first daytime vs. 24 (range, 19-55), 23 (range, 18-57), and 19 (range, 15-53) during the second day period and first and second night period, respectively (p = 0.08). Ineffective effort indexes per hour also were similar across the four periods.
Discussion
In a study where spontaneously breathing patients were conscious and under mechanical ventilation, proportions of sleep fragmentation sleep architecture and sleep quality were positively influenced by NAVA. In the PSV mode, a low percentage of REM sleep and a high degree of fragmentation were present. NAVA showed a normal percentage of REM sleep with an important decrease in fragmentation.
Less than 15% of the sleep fragmentations in the PSV mode were attributed to apneas and ineffective efforts, whereas in NAVA, no asynchrony (no apnea and no ineffective patient efforts) were recorded. Environmental noise is responsible for 18% of the arousals and awakenings in PSV compared with 21% in NAVA, respectively.
We observed results similar to the Cabello et al. [19] study concerning the rate of fragmentation, the number of central apneas, and the number of ineffective patient efforts during PSV. Another similar finding concerned the increased percentage of REM sleep during the second nighttime period recordings. However, one major difference between our study and the Cabello study is that they did not allocate an even distribution for each of the study periods and ventilatory strategies. Also, they did not allow washout periods between the ventilatory modes, which could possibly contaminate the recordings at the beginning of the next study period. Detecting asynchronies also was different; they used the airway pressure-flow signal and the thoracoabdominal plethysmography, whereas we observed the EAdi signal.
Parthasarathy and Tobin [15] found a lower rate of sleep fragmentation during ACV compared with PSV. This was explained by the central apneas induced by over-assistance during PSV. In fact, tidal volume was much greater during PSV compared with ACV. This was validated by the addition of a dead space to the 11 patients showing central apneas, which significantly decreased the number of apneas.
In the Toublanc et al. study [27], no difference was found in terms of quantity, quality of sleep, and in terms of arousal index between the AC and a low level of PSV assistance for the whole night. Toublanc et al. found that ACV was superior in terms of percentage of slow-wave sleep but not during REM sleep [27]. It is very difficult to compare the results for PSV because of a lack of information on expiratory triggering with Evita 4, number of asynchronies (tidal volume, respiratory rate, and minute ventilation). In the Toublanc study, the majority of patients were affected with COPD and pressure support was adjusted to 6 cmH2O. According to Brochard et al., it is suggested that for COPD patients, the pressure support needed to overcome resistance imposed by the endotracheal tube is higher than non-COPD patients: 12 ± 1.9 vs 5.7 ± 1.5 cmH2O respectively [28]. In the Leleu et al. study, pressure support must be superior to 6 cmH2O, particularly in COPD if the intention is to compensate work of breathing imposed by the endotracheal tube, ventilator circuit, and patient effort to trigger the demand valve during pressure support [29]. A low-pressure support only allows for partial relieve of imposed work of breathing without modifying the work necessary to trigger the demand valve. In the Toublanc study, pressure support set too low in COPD patients resulted in an increase in imposed work of breathing, which can be accounted for in the decrease in SWS and REM.
The Toublanc study offers no information on expiratory triggering, which is somewhat important in COPD patients. Tassaux et al. recently have evaluated the positive impact of shortening inspiratory time in PSV on patient-ventilator asynchronies and the work of breathing in COPD patients. This study also demonstrated that the increase in expiratory trigger up to 70% of peak flow improved synchrony and decreased ineffective efforts without modifying work of breathing or minute ventilation [30].
Bosma et al. evaluated the impact on sleep with other modes of ventilation, such as the proportional assist ventilation (PAV). The objective of PAV, such as NAVA, is to improve patient ventilator synchrony by delivering ventilator assist proportional to patient effort. The study by Bosma et al. shows an improvement in the quality of sleep using PAV compared with PSV during one night sleep [31]. There are similarities between the Bosma study and ours. More specifically, PAV appeared superior to PSV in terms of decrease in arousals, improvement in sleep quality, decrease in amounts of arousals, awakenings per hour, and improved SWS and REM. With NAVA, we observed a decrease in tidal volume by up to 15% during REM sleep, which increased end-tidal CO2 by approximately 4 mmHg. Bosma et al. observed a tidal volume slightly more elevated in PSV compared with PAV (despite similar offloading of the work of breathing), resulting in a higher morning PaCO2 with PAV attributed to lower tidal volume and minute ventilation [31], thus offering perhaps a protection against central apneas. Finally, fewer patient-ventilator asynchronies were observed with PAV with fewer awakenings per hour [31].
Contrary to NAVA, PAV cannot eliminate wasted or ineffective efforts. There was a nonstatistically significant difference in ineffective triggering during inspiration; 19.6 n/hr for PSV vs. 11.6 n/hr for PAV [31]. According to Thille et al. ineffective efforts and double triggering are among the most frequent asynchronies: 85 and 13% respectively [32], which is somewhat contradictory to Bosma et al. who identify auto triggering as the most frequent asynchrony in PSV.
We observed that the absence of central apnea and ineffective efforts in NAVA do not totally explain the great improvement in the SWS and REM sleep. This improvement may be explained in part by a microanalysis of the sleep architecture. The microanalysis suggests an over-assistance with PSV during the N-REM stages, because 100% of the fragmentations in PSV occurred during this stage. The tidal volume decrease in NAVA follows the respiratory physiologic changes during sleep, whereas in PSV we find a tidal volume oscillatory behavior due to constant inspiratory efforts, independent of the sleep stage and produces sequential over-assistance during N-REM sleep leading to a decrease in end-tidal CO2. It is our assumption that improvement of the slow-wave sleep and REM is most probably explained by better patient comfort through better neuromechanical coupling.
During sleep, the respiratory accessory muscles (intercostals, scalene, and abdominals) decrease their muscle tone and the mechanical response of the diaphragm is, in part, spent in the production of a mechanical distortion of the chest wall, secondary to a lack of synchronization between diaphragmatic contraction and the accessory muscles. NAVA improves this mechanical distortion, whereas PSV worsens this distortion by a tidal volume oscillation (overshoot) during sleep, with a constant patient effort. Patient comfort is not only directly related to inefficient efforts and central apneas; the microanalysis showed that during N-REM sleep in PSV, the trigger delay increased during stage 1 versus during stage 3 and 4. The expiratory trigger increased in PSV during stage 1 and stages 3 and 4, respectively. In NAVA, the trigger delay remained stable during stage 1 and during stages 3 and 4. The expiratory trigger also remained stable in NAVA, during stage 1 and during stages 3 and 4. NAVA allows optimizing the neuromechanical coupling and therefore patient-ventilator synchrony [33] and allows for optimized adequacy between ventilatory load and patient breathing ability, thereby providing beneficial effects on sleep in ICU patients. It appeared to us that the EAdi tracing is much more efficient than flow and pressure tracings to detect asynchronies.
Our study has some limitations; one is the open space between patients. This study included only 14 patients, which could favor the possibility of a type II error. Patients' heterogeneity implies that patients required bedside care, such as suctioning or other care, which could perhaps influence sleep fragmentation. The study by Cabello found that suctioning was associated with < 1% arousals and awakenings [19]. The choice for a 15-second interval between asynchrony and the occurrence of arousal was chosen based on one previous study on the same topic [19]. Literature on this specific time interval to choose is very scarce. In one study, it was shown that the breathing response to a complete airway occlusion was 20.4 ± 2.3 sec during NREM and 6.2 ± 1.2 sec during REM [34]. The choice of a 15-second interval seems very reasonable but may need further investigation.
In a sleep laboratory, it is a lot easier to control the baseline ambient noise level. In a clinical environment, such as an ICU, we recorded the average baseline ambient noise level and evaluated arousals from this baseline to a peak noise level ≥10 dB above ambient noise level. There is therefore a potential for statistical inaccuracies.
The fact that we stopped sedation 24 hours before beginning the study does not imply an absence of cumulative sedation. However, every patient had a Ramsay Score of 2 or less and a Glasgow Score of 11 (the maximum score for an intubated patient).
Conclusions
The ventilatory mode NAVA improves the quality of sleep by increasing the slow-wave sleep and REM and by decreasing fragmentation. NAVA improves patient comfort through better neuromechanical coupling during N-REM sleep, by a shorter trigger delay, and more efficient expiratory triggering. To minimize sleep fragmentation, optimal setting of pressure support level and expiratory trigger are paramount in PSV. However, proportional assistance modes of ventilation according to patient inspiratory effort, such as NAVA, appear to be a better choice to minimize sleep fragmentation.
References
Drouot X, Cabello B, d'Ortho MP, et al.: Sleep in the intensive care unit. Sleep Med Rev 2008, 12: 391–403. 10.1016/j.smrv.2007.11.004
Friese RS: Sleep and recovery from illness and injury: a review of theory, current practice, and future directions. Crit Care Med 2008, 36: 697–705. 10.1097/CCM.0B013E3181643F29
Gabor JY, Cooper AB, Crombach SA, et al.: Contribution of the Intensive Care Unit Environment to sleep disruption in mechanically ventilated patients and healthy subjects. Am J Respir Crit Care Med 2003, 167: 708–715. 10.1164/rccm.2201090
Bryant PA, Trinde J, Curtis N: Sick and tired: Does sleep have a vital role in the immune system? Nat Rev Immunol 2004, 4: 457–467. 10.1038/nri1369
Valente M, Placidi F, Oliveira AJ, et al.: Sleep organization pattern as a prognostic marker at the subacute stage of posttraumatic coma. Clin Neurophysiol 2002, 113: 1798–1805. 10.1016/S1388-2457(02)00218-3
Helton MC, Gordon SH, Nunnery SL: The correlation between sleep deprivation and the intensive care unit syndrome. Heart Lung 1980, 9: 464–468.
Chen HI, Tang YR: Sleep loss impairs inspiratory muscle endurance. Am Rev Respir Dis 1989, 140: 907–909.
Andrews P, Azoulay E, Antonelli M, et al.: Year in review in intensive care medicine 2004. Part I. Respiratory failure, infection and sepsis. Intensive Care Med 2005, 31: 28–40. 10.1007/s00134-004-2529-5
Andrews P, Azoulay E, Antonelli M, et al.: Year in review in intensive care medicine 2005. Part II. Acute respiratory failure and acute lung injury, ventilation, hemodynamics, education renal failure. Intensive Care Med 2006, 32: 207–216. 10.1007/s00134-005-0027-z
Aaron JN, Carlisle CC, Carskadon MA, et al.: Environmental noise as a cause of sleep disruption in an intermediate respiratory care unit. Sleep 1996, 19: 707–710.
Freedman NS, Gazendam J, Levan L, et al.: Abnormal sleep/wake cycles and the effect of environmental noise on sleep disruption in the intensive care unit. Am J Respir Med 2001, 163: 451–457.
Hilton BA: Quantity and quality of patients' sleep and sleep-disturbing factors in a respiratory intensive care unit. J Adv Nurs 1976, 1: 453–468. 10.1111/j.1365-2648.1976.tb00932.x
Cooper AB, Thornley KS, Young GB, et al.: Sleep in critically ill patients requiring mechanical ventilation. Chest 2000, 117: 809–818. 10.1378/chest.117.3.809
Freedman NS, Kotzer N, Schwab RJ: Patient perception of sleep quality and etiology of sleep disruption in the intensive care unit. Am J Respir Crit Care Med 1990, 159: 1155–1162.
Parthasarathy S, Tobin MJ: Effect of ventilator mode on sleep quality in critically ill patients. Am J Respir Crit Care Med 2002, 166: 1423–1429. 10.1164/rccm.200209-999OC
Nakayama H, Smith CA, Rodman JR, et al.: Effect of ventilatory drive on carbon dioxide sensitivity below eupnea during sleep. Am J Respir Crit Care Med 2002, 165: 1251–1260. 10.1164/rccm.2110041
Thille AW, Rodriguez P, Cabello B, et al.: Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med 2006, 32: 1515–1522. 10.1007/s00134-006-0301-8
Fanfulla F, Delmastro M, Berardinelli A, et al.: Effects of different ventilator settings on sleep and inspiratory effort in patients with neuromuscular disease. Am J Respir Crit Care Med 2005, 172: 619–624. 10.1164/rccm.200406-694OC
Cabello B, Thille AW, Drouot X, et al.: Sleep quality in mechanically ventilated patients: comparison of three ventilatory modes. Crit Care Med 2008, 36: 1749–1755. 10.1097/CCM.0b013e3181743f41
Jasper HH: The ten twenty electrode system of the International Federation. Electroencephal Clin Neurophysiol 1958, 10: 371–375.
Rechtschaffen A, Kales A: A manual of standardized terminology, techniques and scoring system for sleep stages of human subjects. Los Angeles, UCLA BIS/BRI; 1968.
Iber C, Ancoli-Israel S, Chesson A, et al.: The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications, 1st ed. Westchester, Illinois. American Academy of Sleep Medicine; 2007.
EEG arousals, Scoring rules and examples: A preliminary report from the Sleep Disorders Atlas Task Force of the American Sleep Disorders Association. Sleep 1992, 15: 173–184.
Force TRoaAAoSMT: Sleep-related breathing disorders in adults: Recommendation for syndrome definition and measurement techniques in clinical research. Sleep 1999, 22: 667–689.
Xie A, Wong B, Phillipson EA, et al.: Interaction of hyperventilation and arousal in the pathogenesis of idiopathic central sleep apnea. Am J respire Crit Care Med 1994, 150: 489–495.
Trinder J, Merson R, Rosenberg JI, et al.: Pathophysiological interactions of ventilation, arousals, and blood pressure oscillations during Cheyne-Stokes respiration in patients with heart failure. Am J Respir Crit Care Med 2000, 162: 808–813.
Toublanc B, Rose D, Glerant JC, et al.: Assist-control ventilation vs. low levels of pressure support ventilation on sleep quality in intubated ICU patients. Intensive Care Med 2007, 33: 1148–1154. 10.1007/s00134-007-0659-2
Brochard L, Pluskwa F, Lemaire F: Improved efficacy of spontaneous breathing with inspiratory pressure support. Am Respir Dis 1987, 136: 411–415. 10.1164/ajrccm/136.2.411
Leleu O, Mayeux I, Journieaux V: Effets de l'adjonction d'une aide inspiratoire de 6 cmH2O sur la consummation en oxygène des muscles respiratoires au cours du sevrage de la ventilation mécanique. Rev Mal Respir 2001, 18: 283–288.
Tassaux D, Gainnier M, Battisti A, et al.: Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 2005, 172: 1283–1289. 10.1164/rccm.200407-880OC
Bosma K, Ferreyra G, Ambrogio C, et al.: Patient-ventilator interaction and sleep in mechanically ventilated patients: pressure support versus proportional assist ventilation. Crit Care Med 2007, 35: 1048–1054. 10.1097/01.CCM.0000260055.64235.7C
Thille AW, Rodriguez P, Cabello B, et al.: Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med 2006, 32: 1515–1522. 10.1007/s00134-006-0301-8
Schmidt M, Demoule A, Crasso C, et al.: Neurally adjusted ventilatory assist increases respiratory variability and complexity in acute respiratory failure. Anesthesiology 2010, 112: 670–681. 10.1097/ALN.0b013e3181cea375
Issa FG, Sullivan CE: Arousal and breathing responses to airway occlusion in healthy sleeping adults. J Appl Physiol 1983, 55: 1113–1119.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no competing interests.
Authors' contributions
SD and PO drafted the manuscript, and PB, JPT, and PA revised the manuscript.
Authors’ original submitted files for images
Below are the links to the authors’ original submitted files for images.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Delisle, S., Ouellet, P., Bellemare, P. et al. Sleep quality in mechanically ventilated patients: comparison between NAVA and PSV modes. Ann. Intensive Care 1, 42 (2011). https://doi.org/10.1186/2110-5820-1-42
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/2110-5820-1-42