Skip to main content

End-tidal carbon dioxide monitoring using a naso-buccal sensor is not appropriate to monitor capnia during non-invasive ventilation

Abstract

Background

In acute respiratory failure, arterial blood gas analysis (ABG) is used to diagnose hypercapnia. Once non-invasive ventilation (NIV) is initiated, ABG should at least be repeated within 1 h to assess PaCO2 response to treatment in order to help detect NIV failure. The main aim of this study was to assess whether measuring end-tidal CO2 (EtCO2) with a dedicated naso-buccal sensor during NIV could predict PaCO2 variation and/or PaCO2 absolute values. The additional aim was to assess whether active or passive prolonged expiratory maneuvers could improve the agreement between expiratory CO2 and PaCO2.

Methods

This is a prospective study in adult patients suffering from acute hypercapnic respiratory failure (PaCO2 ≥ 45 mmHg) treated with NIV. EtCO2 and expiratory CO2 values during active and passive expiratory maneuvers were measured using a dedicated naso-buccal sensor and compared to concomitant PaCO2 values. The agreement between two consecutive values of EtCO2 (delta EtCO2) and two consecutive values of PaCO2 (delta PaCO2) and between PaCO2 and concomitant expiratory CO2 values was assessed using the Bland and Altman method adjusted for the effects of repeated measurements.

Results

Fifty-four datasets from a population of 11 patients (8 COPD and 3 non-COPD patients), were included in the analysis. PaCO2 values ranged from 39 to 80 mmHg, and EtCO2 from 12 to 68 mmHg. In the observed agreement between delta EtCO2 and deltaPaCO2, bias was −0.3 mmHg, and limits of agreement were −17.8 and 17.2 mmHg. In agreement between PaCO2 and EtCO2, bias was 14.7 mmHg, and limits of agreement were −6.6 and 36.1 mmHg. Adding active and passive expiration maneuvers did not improve PaCO2 prediction.

Conclusions

During NIV delivered for acute hypercapnic respiratory failure, measuring EtCO2 using a dedicating naso-buccal sensor was inaccurate to predict both PaCO2 and PaCO2 variations over time. Active and passive expiration maneuvers did not improve PaCO2 prediction.

Trial registration

ClinicalTrials.gov: NCT01489150.

Background

Non-invasive ventilation (NIV) is widely used [1] in emergency rooms, in intensive and intermediate care units, and in recovery rooms to treat de novo and, even if it is more debatable [2,3], postextubation hypercapnic respiratory failure. Arterial blood gas analysis (ABG) is usually performed to diagnose hypercapnia and should at least be repeated within 1 h after NIV initiation to assess PaCO2 response to treatment [1]. However, as follow-up ABG requires a new arterial puncture in patients not previously equipped with an arterial line, this exam is often postponed with the risk of delaying NIV failure diagnosis and intubation, a condition previously associated with poor outcome [4]. Only a reliable non-invasive monitoring of the course of PaCO2 during NIV could avoid such a delay and help optimizing ventilator settings. End-tidal CO2 (EtCO2) monitoring is easy to perform and widely used during anesthesia to assess the adequacy of delivered minute ventilation without performing repetitive ABG [5,6]. Using capnometry to monitor capnia in non-intubated patients during NIV is much more challenging. Indeed, during NIV, gas leak occurs in the respiratory ‘circuit’ and conceivably, in this situation, only gas sampling directly at the level of the patient’s airways can reflect true expiratory gas.

As new specialized naso-buccal EtCO2 sensors have recently been developed to collect expired gas directly at the airway opening, there is now an opportunity to use capnometry to monitor capnia during NIV. The main aim of this study was to assess the ability of a dedicated EtCO2 naso-buccal sensor to predict PaCO2 variations and/or PaCO2 absolute values in hypercapnic patients during NIV. The second aim of the study was to assess whether active or passive prolonged expiratory maneuvers could improve the agreement between expiratory CO2 and PaCO2.

Methods

A prospective pilot study was conducted in our medico-surgical ICU in Lausanne, Switzerland. The hospital ethics committee (Human Research Ethics Committee of Lausanne, Switzerland) approved the study protocol, and written informed consent was obtained before inclusion in the study. In the absence of published data reporting the use of a naso-buccal sensor to measure EtCO2 in acutely ill patients undergoing NIV, no power computation could be performed.

Patients

Non-intubated patients suffering from hypercapnic (PaCO2 ≥ 45 mmHg) acute respiratory failure, hospitalized in the ICU, equipped with an arterial line and requiring NIV could be included in the study if they had no major hemodynamic instability, no facial lesion preventing the use of the naso-buccal sensor, and no cognitive disability or psychiatric disease liable to interfere with NIV. To note, as only patients admitted in the ICU and already equipped with an arterial line could be included in the study, the NIV treatment monitored in the study was usually not the first NIV treatment delivered to the patients.

Study protocol and measurements

Upon inclusion, the patient was equipped with the Smart CapnoLine® naso-buccal sensor (Figure 1) designed to collect expiratory gas immediately at the airway opening both at the nose and mouth levels connected to the Capnostream 20 monitor® (Oridion Medical Ltd, Jerusalem, Israël). To perform the measurement, a sample of gas is transmitted from the patient to a micro-cell of 15 μl located in the monitor (sidestream capnography system). A sample of gas of 50 ml/min is needed for the measurement. The measurement is performed by non-dispersive infrared spectroscopy. For each respiratory cycle, the capnogram is displayed on the Capnostream 20 monitor®. For each EtCO2 value recorded, the investigator checked the good quality of the capnogram displayed on the screen. The value of one respiratory cycle was recorded at each measurement time.

Figure 1
figure 1

Naso-buccal sensor. Illustration of the naso-buccal sensor used in this study. This device is designed to collect expiratory gas immediately at the airway opening both at the nose and mouth levels.

ABG and the corresponding EtCO2 value displayed by the monitor were recorded as baseline values. NIV treatment was then initiated using an hermetic naso-buccal mask (Vygon Large®, Ecouen, France) held in place using a dedicated strap and a single-limb NIV ventilator (V60®, Respironics Philips, Amsterdam, Netherland). Calibrated intentional leakage to allow CO2 expiration was created in the respiratory circuit using the dedicated whisper swivel (Whisper swivel®, Respironics Philips, Amsterdam, Netherland). A flow sensor (Hamilton, Bonaduz, Switzerland) was placed between the patient and the whisper swivel and connected to an analog-to-digital converter (MP100, Biopac, Systems, Goleta, CA, USA) to continuously record the flow-time curve. The respiratory circuit with the additional flow sensor is schematized in Figure 2. ABG and EtCO2 values were recorded at 15, 30, 45, and 60 min after the initiation of NIV. At times corresponding to each PaCO2 and EtCO2 measurements, insufflated volumes were measured offline for ten consecutive respiratory cycles (by integration of the inspiratory flow-time curve recorded by the flow sensor placed between the patient and the whisper swivel) and the mean value was computed. Respiratory rate and delivered minute ventilation were also computed.

Figure 2
figure 2

Respiratory circuit. Illustration of the respiratory circuit used in the study. From the patient to the ventilator, the circuit consists of an hermetic nasobuccal mask, the dedicated proximal flow sensor of the V60 ventilator, the additional flow sensor inserted to record insufflated and exuflated flow-time curves, the dedicated whisper swivel to create a calibrated intentional leak to avoid CO2 rebreathing, and the ventilator single-limb pipe.

At 30 and 60 min after the beginning of NIV, the patient performed upon request a voluntary slow and maximal expiration. In brief, the patients were asked to slowly empty their lungs as much and for as long as possible. The expired CO2 value displayed at the end of this active expiration maneuver was recorded. A passive expiratory maneuver was then performed with the help of an experienced respiratory therapist (bilateral chest compression during slow expiration), and the corresponding expired CO2 value was recorded. The naso-buccal mask was not removed during the maximal expiratory maneuvers meaning that the patient expired through the nasobuccal sensor and the ventilator circuit and thus against the set PEEP. The backup safety respiratory frequency of the ventilator was set at 6 by minute to allow expiratory maneuvers of 10 s.

Calculations and statistics

To assess PaCO2 variations, the differences between two consecutive PaCO2 (delta PaCO2) values were computed for each patient between the initial value and the 15-min value, between the 15- and 30-min values, between the 30- and 45-min values, and finally between the 45- and 60-min values. Delta EtCO2 were computed to assess EtCO2 variations according to the same procedure.

The PaCO2-EtCO2 gradient (Pa-E′CO2) was computed for each patient with the pair of values recorded at the beginning of the NIV session and at 15, 30, 45 and 60 min after the initiation of NIV. The number of Pa-E′CO2 values of more than ±10 mmHg was reported. The ratio of this number over the total number of measurements represents the proportion of clinically unacceptable EtCO2 values. The treshold of 10 mmHg to consider Pa-E′CO2 as clinically acceptable or not was an arbitrary choice.

All statistical analyses were performed using MedCalc Statistical Software version 12.7.2 (MedCalc Software, Ostend, Belgium). Considering the small number of included patients, non-normal distribution of the results was assumed. All results are given as median [25th and 75th percentile].

The agreement between delta PaCO2 and delta EtCO2 was assessed by the Bland and Altman method adjusted for the effect of repeated measurements. The differences between each deltaPaCO2 and deltaEtCO2 values were also computed. The percentage of differences higher than 5 mmHg was reported as they were arbitrarily considered as clinically unacceptable values.

Agreement between PaCO2 and EtCO2 absolute values was assessed using the Bland and Altman method adjusted for the effects of repeated measurements. Expiratory CO2 to PaCO2 agreement for values obtained after active and passive complete expirations was also computed with the Bland and Altman method adjusted for the effects of repeated measurements. The gradient between expiratory CO2 and PaCO2 was computed with the values obtained after active and passive complete expirations respectively. Clinically unacceptable values were arbitrarily defined as values above 10 mmHg. The proportions of clinically unacceptable gradients recorded were compared between normal expiration, active complete expiration, and passive complete expiration by chi-square test. p < 0.05 was considered as statistically significant.

Results

The whole 45-min protocol could be applied to ten patients. In one patient (patient number 4), the NIV treatment had to be interrupted after 45 min because of intolerance. In this patient, the second set of active and passive expiratory manoeuvers was performed after 45 min instead of 1 h, immediately before stopping NIV. Overall, 54-paired data sets of PaCO2 and EtCO2 from 11 patients (seven men/four women) could be recorded and were included in the analysis. Patients’ demographic and clinical data are given in Table 1. Among the 11 included patients, eight patients had chronic obstructive pulmonary disease (COPD) of various severity (Table 1). Median age was 68 [62 and 77] years old and median SAPS II score was 43 [34 and 44]. Initial blood gas analysis, respiratory rate, inspired fraction of oxygen (FIO2), PaO2/FIO2 ratio, and initial ventilator settings during NIV are mentioned in Table 2.

Table 1 Patient’s characteristics and clinical information.
Table 2 Respiratory rate, blood gas analysis at inclusion, and main initial ventilator settings

During the study period, PaCO2 ranged from 39 to 80 mmHg, and EtCO2 from 12 to 68 mmHg. At the time of the measurements, delivered inspiratory volume was 724 [597–896] ml and delivered minute ventilation was 18.6 [14.0-22.7] l/min. When assessing the agreement between EtCO2 and PaCO2 gradients between two consecutive measurements, 43 paired data sets could be analyzed. The bias was −0.3 mmHg and the limits of agreement were −17.8 and +17.2 mmHg. The Bland and Altman graphic representation is displayed in Figure 3. Sixteen of 43 differences (37%) between delta PaCO2 and delta EtCO2 were higher than 5 mmHg.

Figure 3
figure 3

Bland-Altman plot of agreement between delta PaCO 2 and delta EtCO 2 . Bland-Altman plot of agreement between delta PaCO2 and delta EtCO2. PaCO2, CO2 partial pressure in arterial blood; EtCO2, end-tidal CO2; circle markers, COPD patients values; square markers, non-COPD patients values; COPD, chronic obstructive pulmonary disease. The horizontal lines represent the bias and the upper and lower limits of agreement.

When assessing agreement between PaCO2 and EtCO2 absolute values, bias was 14.7 mmHg and the limits of agreement were −6.6 and 36.1 mmHg (Figure 4). The Bland and Altman graphic representation is displayed in Figure 4 both for COPD patients and non-COPD patients. Pa-E′CO2 was 12.4 [8.6-20.2] mmHg in median but very high values were documented in some patients (maximal value of 42.7 mmHg) and non-physiologic slightly negative values were observed in one patient (Figure 5). The number of clinically unacceptable values for Pa-E′CO2 was 35/54 (65%).

Figure 4
figure 4

Bland-Altman plot of agreement between PaCO 2 and EtCO 2 . Bland-Altman plot of agreement between PaCO2 and EtCO2. PaCO2, CO2 partial pressure in arterial blood; EtCO2, end-tidal CO2; circle markers, COPD patients values; square markers, non-COPD patients values; COPD, chronic obstructive pulmonary disease. The horizontal lines represent the bias and the upper and lower limits of agreement.

Figure 5
figure 5

Evolution over time of PaCO 2 -EtCO 2 gradient for all the patients. This figure shows the evolution over time of PaCO2-EtCO2 gradient for all the patients. PaCO2, CO2 partial pressure in arterial blood; EtCO2, end-tidal CO2. Patient numbers 1, 3, 4, 5, 7, 9, 10, and 11 are COPD patients.

When we compared agreements between PaCO2, concomitant EtCO2, and expired CO2 after active and passive expiration maneuvers, we had 22-paired data available for each comparison. The bias was respectively 15.7, 9.9, and 9.8 mmHg. Bland-Altmann plots for active and passive expiration maneuvers are displayed in Figure 6A,B respectively. The number of clinically unacceptable gradient values was not different between the three measurements (respectively, 13 (60%), 9 (41%), and 9 (41%), p = 0.37).

Figure 6
figure 6

Bland-Altman plot of agreement between PaCO 2 and expired CO 2 after active and passive maximal expiration maneuvers. A Bland-Altman plot of agreement between PaCO2 and expired CO2 after active maximal expiration maneuver (ExPA). B Bland-Altman plot of agreement between PaCO2 and Expired CO2 after passive maximal expiration maneuver (ExPP). PaCO2, CO2 partial pressure in arterial blood. In both figures, the horizontal lines represent the bias and the upper and lower limits of agreement.

Discussion

Our results show that, in patients suffering from hypercapnic acute respiratory failure, measuring EtCO2 by a dedicated naso-buccal sensor during NIV was inaccurate to predict either PaCO2 variation over time or the absolute PaCO2 value. Adding complete passive or active expiratory maneuvers to expiratory CO2 measurements did not significantly improve the reliability of PaCO2 prediction.

Before discussing the results in more details, we must acknowledge the following limitations of our study. First, only a small number of patients were included. However, a high number of paired EtCO2 and PaCO2 could be analyzed. As the correlation was poor with very high limits of agreements, it is unlikely that increasing the number of patients would have significantly modified the results. Second, this study used a specific system to measure EtCO2 and we cannot exclude that using another device could have yielded different results. Third, only one EtCO2 value was recorded at each time. Even if the quality of the corresponding capnogram was carefully checked, we cannot exclude that averaging the values of several respiratory cycles could have provided slightly different results. However, as airway resistance usually not varies between one breath and the following, this effect, if present, should be minor. Fourth, using another patient-ventilator interface or other ventilators, e.g., ICU ventilators equipped with inspiro-expiratory circuits, might also lead to different results. Fifth, during the active and passive complete expiration maneuvers, some patients could potentially not have emptied their lungs enough to reach the residual volume because of maneuver intolerance or because they had to expire through the breathing circuit against the set PEEP. Thus, expired CO2 values might not truly reflect expired CO2 at residual lung volume. Finally, we cannot exclude that different results could have been found if we had measured EtCO2 after stopping NIV treatment. However, as, in clinical practice, it can be difficult or even dangerous to interrupt NIV treatment in patients suffering from acute respiratory failure, we did not test this alternative approach.

EtCO2 has been efficiently used for decades in intubated anesthetized patients [7] to monitor PaCO2 and ventilation, although many limitations have been recognized, particularly for patients suffering from chronic respiratory diseases (increased VD/VT ratio [8], airflow limitation) or hemodynamic instability leading to ventilation-perfusion mismatches [7,9]. Nasal EtCO2 has been successfully used to monitor normocapnic patients with almost healthy lungs undergoing regional anesthesia or recovering from general anesthesia [10]. In line with the results of the present study, two studies performed in spontaneously breathing patients suffering from acute respiratory failure found poor agreement between EtCO2 and PaCO2 values [11,12]. Oppositely, in more stable and tracheotomized patients, EtCO2 values were closer to PaCO2 values [13].

In contrast to our results (see Figure 4), in this last study [13], the agreement between EtCO2 and PaCO2 was better in non-COPD patients than in those suffering from COPD. This last point suggests that during NIV, physiopathological reasons probably do not explain by themselves the poor performances of EtCO2 measurement. A possible explanation for the poor agreement we observed between EtCO2 and PaCO2 during NIV could be the presence of a high airflow and of significant and often variable leaks during NIV that may have caused sampled expiratory gas dilution.

To try to overcome the expected limitation of EtCO2 measurement to assess PaCO2 absolute values and based on the assumption that, in the absence of major haemodynamic instability and of bronchodilatator administration, Pa-E′CO2, even if often unpredictable, might be sufficiently constant over an hour in a given patient to enable the tracking of PaCO2 evolution, we assessed the time evolution of EtCO2 and PaCO2. This approach clearly reduced the bias, but the wide limits of agreement preclude its clinical use. Of course, we cannot exclude that physiological reasons, as alveolar recruitment occurring during NIV, could have decreased the VD/VT ratio and contibutated to the poor performance of EtCO2 variations to assess PaCO2 variations during NIV. However in this situation, EtCO2 values would have been closer to PaCO2 values at the end of the 1-h NIV treatment, which was not the case.

To try to better assess PaCO2, we also attempted to sample gas closer to the alveolar compartment by measuring expiratory CO2 at the end of a ‘complete’ expiration (either active or passive) [14] but this approach was also disappointing. Again, this observation contrasts with a study on stable tracheostomized patients [13] and underlines that performing reliable complete expiration maneuvers in acutely ill patients is very difficult.

The present study suggests that other technologies should be considered to non-invasively assess PaCO2 and PaCO2 over time during NIV. Even if the reliability of using transcutaneous CO2 monitoring to assess PaCO2 in case of acute respiratory failure is still contoversial [15,16], recent technological improvements in the transcutaneous CO2 monitoring technology suggest that this technique could be of interest to monitor PaCO2 during NIV. This hypothesis, however, should be formally explored prospectively.

Conclusions

When a naso-buccal sensor is used, major variations of Pa-E′CO2 along time and poor limits of agreements between EtCO2 and PaCO2 preclude the use of EtCO2 measurement to predict PaCO2 or its variation over time during NIV delivered for acute hypercapnic respiratory failure. Adding complete expiration maneuvers, whether passive or active did not improve PaCO2 prediction using EtCO2 during NIV. The optimal approach to non-invasively monitor PaCO2 during NIV in patients with acute hypercapnic respiratory failure remains to be determined.

References

  1. Nava S, Hill N. Non-invasive ventilation in acute respiratory failure. Lancet. 2009;374:250–9.

    Article  PubMed  Google Scholar 

  2. Ferrer M, Esquinas A, Arancibia F, Bauer TT, Gonzalez G, Carrillo A, et al. Noninvasive ventilation during persistent weaning failure: a randomized controlled trial. Am J Respir Crit Care Med. 2003;168:70–6.

    Article  PubMed  Google Scholar 

  3. Keenan SP, Powers C, McCormack DG, Block G. Noninvasive positive-pressure ventilation for postextubation respiratory distress: a randomized controlled trial. JAMA. 2002;287:3238–44.

    Article  PubMed  Google Scholar 

  4. Esteban A, Frutos-Vivar F, Ferguson ND, Arabi Y, Apezteguia C, Gonzalez M, et al. Noninvasive positive-pressure ventilation for respiratory failure after extubation. N Engl J Med. 2004;350:2452–60.

    Article  CAS  PubMed  Google Scholar 

  5. Takki S, Aromaa U, Kauste A. The validity and usefulness of the end-tidal pCO 2 during anaesthesia. Ann Clin Res. 1972;4:278–84.

    CAS  PubMed  Google Scholar 

  6. Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg. 1981;60:508–12.

    Article  CAS  PubMed  Google Scholar 

  7. Moon RE, Camporesi EM. Respiratory monitoring. In: Miller DR, editor. Miller’s anesthesia. Volume 4th edition. 4th ed. New York: Churchill Livingtone; 1994. p. 1253–91.

    Google Scholar 

  8. Hoffman RA, Krieger BP, Kramer MR, Segel S, Bizousky F, Gazeroglu H, et al. End-tidal carbon dioxide in critically ill patients during changes in mechanical ventilation. Am Rev Respir Dis. 1989;140:1265–8.

    Article  CAS  PubMed  Google Scholar 

  9. Fletcher R, Jonson B, Cumming G, Brew J. The concept of deadspace with special reference to the single breath test for carbon dioxide. Br J Anaesth. 1981;53:77–88.

    Article  CAS  PubMed  Google Scholar 

  10. Raheem MS, Wahba OM. A nasal catheter for the measurement of end-tidal carbon dioxide in spontaneously breathing patients: a preliminary evaluation. Anesth Analg. 2010;110:1039–42.

    Article  PubMed  Google Scholar 

  11. Jabre P, Jacob L, Auger H, Jaulin C, Monribot M, Aurore A, et al. Capnography monitoring in nonintubated patients with respiratory distress. Am J Emerg Med. 2009;27:1056–9.

    Article  PubMed  Google Scholar 

  12. Delerme S, Freund Y, Renault R, Devilliers C, Castro S, Chopin S, et al. Concordance between capnography and capnia in adults admitted for acute dyspnea in an ED. Am J Emerg Med. 2010;28:711–4.

    Article  PubMed  Google Scholar 

  13. Johnson DC, Batool S, Dalbec R. Transcutaneous carbon dioxide pressure monitoring in a specialized weaning unit. Respir Care. 2008;53:1042–7.

    PubMed  Google Scholar 

  14. Takano Y, Sakamoto O, Kiyofuji C, Ito K. A comparison of the end-tidal CO2 measured by portable capnometer and the arterial PCO2 in spontaneously breathing patients. Respir Med. 2003;97:476–81.

    Article  CAS  PubMed  Google Scholar 

  15. Gancel PE, Roupie E, Guittet L, Laplume S, Terzi N. Accuracy of a transcutaneous carbon dioxide pressure monitoring device in emergency room patients with acute respiratory failure. Intensive Care Med. 2011;37:348–51.

    Article  PubMed  Google Scholar 

  16. Kelly AM, Klim S. Agreement between arterial and transcutaneous PCO2 in patients undergoing non-invasive ventilation. Respir Med. 2011;105:226–9.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

The EtCO2 recording system and the dedicated nasal canulas were kindly provided free of charge by Oridion Medical Ltd., Jerusalem, Israël.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lise Piquilloud.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

LP participated in the design of the study, conducted the study, contributed to statistical analysis, and drafted the manuscript. DT participated in the design of the study, collected the data, and helped with the data analysis and presentation. PJ participated in the design of the study and extensively revised the manuscript. JPR participated in the design of the study, performed the statistical analysis, and extensively revised the manuscript. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Piquilloud, L., Thevoz, D., Jolliet, P. et al. End-tidal carbon dioxide monitoring using a naso-buccal sensor is not appropriate to monitor capnia during non-invasive ventilation. Ann. Intensive Care 5, 2 (2015). https://doi.org/10.1186/s13613-014-0042-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13613-014-0042-8

Keywords