Methodological considerations in personalized methods for PEEP optimization with electrical impedance tomography

Dear editor

We have read with interest the recently published article by Pavlovsky and colleagues [1]. The authors examined patients with ARDS during a decremental PEEP trial performed with steps of 3 cmH2O from 20 to 5 cmH2O. Airway and esophageal pressures as well as electrical impedance tomography (EIT) data were recorded simultaneously. Four different approaches were then applied to identify the ’optimum PEEP’ value from the recordings of each patient by determining those PEEP levels of the PEEP trial at which (1) the plateau pressure was 28–30 cmH2O, (2) the positive expiratory transpulmonary pressure reached its minimum (’positive PLE’), (3) the center of ventilation (CoV) was closest to 0.5 and (4) the relative lung overdistension and collapse curves crossed (’crossing point’). The latter two values were derived from the EIT data. With the exception of the ’crossing point’ and ’positive PLE’, the calculated ’optimum PEEP’ values varied significantly from each other, and none of the values was correlated to the recruitment-to-inflation ratio (R/Iest), as calculated from the change in end-expiratory lung impedance between PEEP of 5 and 15 cmH2O.

In physiological studies investigating different parameters for individualization of PEEP settings, correct understanding and interpretation of the examined parameters is of utmost importance. Regarding the R/Iest parameter, a previous study suggested that its absolute value does not correlate with actual lung recruitment because this parameter cannot distinguish atelectasis and overdistension [2]. In the current study by Pavlovsky and colleagues [1], the ’optimum PEEP’ levels determined with the four abovementioned methods were not associated with R/Iest, leading the authors to the conclusion that ”optimal PEEP levels proposed by these four methods were not associated with recruitability”. This conclusion may be inaccurate because R/Iest ignores overdistension and can be misleading. Since R/Iest is certainly not the gold standard for PEEP setting, as implicitly presented by the authors, it would be interesting to see if any of the other parameters were correlated with respect to ’optimum PEEP’ (e.g., was there a positive correlation between the ’positive PLE’ and ’crossing point’ PEEP values? ).

Besides these general methodological considerations, we would like to point out some important inaccuracies in the assessment of the EIT-derived CoV parameter. The top three panels of Fig. 1 show the individual patient values of CoV, relative overdistension and collapse. Surprisingly, the highest CoV values were identified at the lowest PEEP value of 5 cmH2O. The authors wrote in the Methods that CoV was ”the percentage of ventilation reaching the dorsal half of the lung”. The dorsal fraction of ventilation should not be the highest at the lowest PEEP level when derecruitment is most likely to have occurred in the dependent lung regions. The presented CoV values also contradict the findings presented in the top right panel showing the highest percentage of lung collapse at the lowest PEEP. The presented results would only make sense if the patients were examined in the prone position, however, the authors wrote that the patients were studied in the semi-recumbent position.

In addition to these physiologically implausible results, the authors also disregarded that CoV and the dorsal fraction of ventilation are two separate, not mutually interchangeable EIT parameters [3]. They are calculated differently and they provide dissimilar information on ventilation distribution. CoV is determined as the weighted geometrical center of ventilation distribution and its location is projected on the ventrodorsal axis of the chest and given in percent of chest diameter. Values lower than 50% imply more ventral and values higher than 50% more dorsal location. The dorsal fraction of ventilation, however, gives just the proportion of ventilation identified in the dorsal half of the analyzed functional EIT image. The Consensus Statement on Chest EIT [4] (which the authors cite in their article´s reference list) provides a clear description of the EIT measures and recommendations for unified use and reporting. If such recommendations are not adopted, the readers may become confused and the users of EIT technology could even draw wrong conclusions on the performance of EIT parameters.

Previous studies indicated that the dorsal fraction of ventilation showed large variation in lung-healthy patients and CoV was mostly located in the ventral part of the chest [5]. Based on these findings, the authors’ decision to define the ’optimum PEEP’ as the value when CoV location was at exactly 50% of the ventrodorsal chest diameter may not be a physiologically sound approach. Indeed, the authors found that CoV reached the value close to 50% of the chest diameter only at a PEEP value of 20 cmH2O. Since the overdistension most probably occurred mainly in the non-dependent ventral lung regions, this effect ’pushed’ the CoV towards the center of the chest. We therefore hypothesize that lower (i.e., more ventrally located) CoV values might be more appropriate for setting PEEP in future studies or clinical use.

Data availability

Not applicable.

Abbreviations

ARDS:

Acute respiratory distress syndrome

CoV:

Center of ventilation

EIT:

Electrical impedance tomography

PLE:

Expiratory transpulmonary pressure

PEEP:

Positive end-expiratory pressure

R/Iest :

Recruitment-to-inflation ratio

References

  1. Pavlovsky B, Desprez C, Richard JC, Fage N, Lesimple A, Chean D, Courtais A, Mauri T, Mercat A, Beloncle F. Bedside personalized methods based on electrical impedance tomography or respiratory mechanics to set PEEP in ARDS and recruitment-to-inflation ratio: a physiologic study. Ann Intensive Care. 2024;14(1):1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Zhao Z, Chang MY, Chu CC, Chang HT, Moller K, Frerichs I, Hsu YL. The absolute value of recruitment-to-inflation ratio does not correlate with the recruited volume. Crit Care. 2023;27(1):246.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Frerichs I, Zhao Z, Becher T. Simple electrical impedance tomography measures for the assessment of ventilation distribution. Am J Respir Crit Care Med. 2020;201(3):386–8.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Frerichs I, Amato MB, van Kaam AH, Tingay DG, Zhao Z, Grychtol B, Bodenstein M, Gagnon H, Bohm SH, Teschner E, Stenqvist O, Mauri T, Torsani V, Camporota L, Schibler A, Wolf GK, Gommers D, Leonhardt S, Adler A. Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group. Thorax. 2017;72(1):83–93.

    Article  PubMed  Google Scholar 

  5. Yang L, Dai M, Cao X, Moller K, Dargvainis M, Frerichs I, Becher T, Fu F, Zhao Z. Regional ventilation distribution in healthy lungs: can reference values be established for electrical impedance tomography parameters? Ann Transl Med. 2021;9(9):789.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

IF: first draft of the manuscript, IF, TB, ZZ: revision of the manuscript, IF, TB, ZZ: final approval of the manuscript.

Corresponding author

Correspondence to Inéz Frerichs.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

IF reports funding from the European Commission (project WELMO, grant no. 825572). TB reports consulting fees from Löwenstein Medical Innovation for institutional funding and lecture fees from Vyaire Medical. ZZ receives a consulting fee from Dräger Medical.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Frerichs, I., Becher, T. & Zhao, Z. Methodological considerations in personalized methods for PEEP optimization with electrical impedance tomography. Ann. Intensive Care 14, 62 (2024). https://doi.org/10.1186/s13613-024-01288-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13613-024-01288-0