Sustained low efficiency dialysis should not be interrupted for performing transpulmonary thermodilution measurements

Abstract

Background

Treatment of multiple organ failure frequently requires enhanced hemodynamic monitoring. When renal replacement is indicated, it remains unclear whether transpulmonary thermodilution (TPTD) measurements are influenced by renal replacement therapy (RRT) and whether RRT should be paused for TPTD measurements. Our aim was therefore to investigate the effect of pausing RRT on TPTD results in two dialysis catheter locations.

Materials and methods

In total, 62 TPTD measurements in 24 patients (APACHE: 32 ± 7 [mean ± standard deviation (SD)]) were performed using the PiCCO™ system (Pulsion, Germany). Patients were treated with sustained low efficiency dialysis (SLED; Genius™ system, Fresenius, Germany) as RRT. Measurements were taken during ongoing hemodialysis (HD, HDO), during paused HD (HDP) and immediately after termination of HD and blood restitution (HDT). Dialysis catheters were placed either in the superior vena cava (SVC, 19 times) or in the inferior vena cava (IVC, 5 times). Statistical analysis was performed to assess the effects of the measurement setting, SLED (blood flow rate) and the catheter location, on cardiac index (CI), global end-diastolic volume index (GEDVI) and extravascular lung water index (EVLWI) as measured by TPTD. Multilevel models were used for the analysis due to the triplicate measurements and due to 12 out of 19 SVC and 2 out of 5 IVC patients having more than one TPTD measured.

Results

CI and GEDVI were significantly higher at time point HDP compared to both HDO and HDT. In contrast, values for EVLWI were lower at HDP when compared to HDO and HDT. These findings were independent of the site of dialysis catheter insertion and blood flow rate.

Conclusions

PiCCO™ measurements assessed at paused SLED significantly deviate from ongoing and terminated SLED. Therefore, the dialysis system should not be paused for measurements. TPTD measurements in patients with PiCCO monitoring seem sufficiently reliable during ongoing SLED as well as after its termination. An effect of dialysis catheter location (SVC vs IVC) and blood flow rate on PiCCO™ measurements could not be shown.

Background

Measurement of key parameters such as the cardiac index (CI) is crucial for appropriate hemodynamic monitoring of critically ill patients. Techniques like indicator dilution are considered as most appropriate and can be obtained either by pulmonary arterial catheter or by transpulmonary thermodilution (TPTD) [1, 2]. Since up to 30% of critically ill patients with multiple organ failure (MOF) develop acute kidney injury (AKI), 10% of them will require renal replacement therapy (RRT) [3]. Thus, simultaneous use of invasive hemodynamic monitoring including pulse contour analyses and TPTD during RRT is common practice. Nowadays, sustained low efficiency dialysis (SLED) for RRT is more frequently used as compared to intermittent RRT because of a better hemodynamic tolerability, but also for economic reasons.

There are concerns about the applicability of indicator dilution techniques including TPTD during RRT [4,5,6,7,8]. As TPTD relies on subtle temperature changes in arterial blood in response to central venous injection of a cold saline bolus, a pulsatile extracorporeal circuit might introduce a measurement bias. In addition, there are other potential confounders of the RRT on thermodilution: loss of indicator in the circuit, changes in blood pump flow and finally an immediate proximity of the central venous catheter (CVC) and dialysis catheter (e.g., both catheters in jugular veins or both catheters in the femoral veins), which might impair TPTD by withdrawing the fluid bolus injected via the CVC via the dialysis catheter [3].

Moreover, there is a controversy regarding the optimal time point for measurement [9]. One current matter of discussion is whether the ideal measurement should be performed during ongoing or at paused RRT. Some authors advocate pausing RRT during measurements [4, 5], while others do not observe any influence of pausing the RRT on TPTD results [10,11,12]. As in most of these studies continuous RRT methods like CVVH, CVVHD and CVVHF were assessed, only a few directly evaluated the influence of SLED on TPTD measurements. Recently, Huber et al. published a systematic evaluation of the influence of connected or disconnected SLED on TPTD [3]. The effect of paused RRT on the TPTD measurements in patients with SLED has, to the best of our knowledge, not yet been systematically evaluated. Therefore, the aim of this study is to provide data about the influence of paused SLED on TPTD measurements.

Materials and methods

Patients

Twenty-four critically ill patients (17 males, 7 females, see Table 1 for group characteristics) with MOF, defined as acute renal failure according to the KDIGO classification in combination with a circulatory failure requiring catecholamine therapy, treated in the medical intensive care units of the Clinic for Cardiology, Pneumology and Internal Intensive Care at the Clinic Munich, Schwabing, and the Department of Clinical Toxicology, Klinikum rechts der Isar, Technical University of Munich, were prospectively enrolled during a 12-month study period. The study protocol was approved by the ethics committee of the Faculty for Medicine of the Technical University of Munich (ethics no. 159/14). The study was performed in accordance with the Declaration of Helsinki and Good Clinical Practice ICH-E6.

Table 1 Patient characteristics at baseline during the first dialysis

Hemodynamic assessment

A 13.5-F dialysis catheter (Niagara 13.5 F × 15/20 cm, C.R. Bard Inc., NJ, USA) was placed in either the SVC or the IVC as confirmed by radiological position monitoring. Blood flow rates varied between 150 and 260 mL/min.

Patients also received a multilumen CVC (Arrow 8.5 F × 20 cm, Teleflex Inc., USA) for infusion of medications with the tip in the SVC.

The cardiac index (CI), global end-diastolic volume index (GEDVI) and extravascular lung water index (EVLWI) were measured for each patient in triplicate in three different settings: (1) measurement while ongoing HD (HDO) (SLED, Genius™ system by Fresenius, Germany); (2) measurement at paused HD (HDP) without disconnection of the system for the duration of the measurements; and (3) measurement performed shortly after RRT was terminated after blood restitution (HDT). For each measurement, a bolus injection of 15 mL cooled normal saline was injected through a CVC placed in the SVC. The triplicate measurements were performed consecutively with the smallest possible delay and immediately after release of each injection, using the same approach and device type [PiCCO-Monitor (PiCCO Pod/Infinity Delta, Dräger, Germany)]. In order to guarantee stable hemodynamic conditions, MAP and heart rate were noted during each measurement series and doses of vasopressors, inotropes and analgosedative agents as well as volume application, and ventilation parameters remained unchanged during measurement, ensuring that the clinical condition of the patient at different measurement time points was in a “steady state.”

Statistical analysis

Statistical analysis was performed using R version 3.1.0 [13]. To assess intra-individual changes in CI, EVLWI and GEDVI between each of the three measurement settings (HDO, HDP, HDT), multilevel (mixed) models were fitted due to the triplicate measurements and measurements for more than one dialysis sessions for 12 out of 19 SVC and 2 out of 5 IVC patients. All measurements of a triplicate were included in the analysis. The models therefore included a patient random effect and a dialysis within patient random effect. Each measurement setting (HDO, HDP and HDT), together with the catheter location, was included as fixed effects. To assess the potential effect of blood flow rates on CI, EVLWI and GEDVI, blood flow rate was also included as a fixed effect in additional analyses. The estimated coefficients from these models are presented together with 95% confidence intervals. The baseline data are presented as mean ± standard deviation (SD). To assess the differences in variability between each of the measurement settings, the SD was calculated for the triplicates at each setting for each dialysis run. Paired t-tests were used to compare the SD between measurement settings. To explore if the first measurement of the triplicate leads to higher variability, mean absolute differences between the first and second, second and third and first and third measurements were calculated for each measurement setting. The coefficient of variation (CV) was calculated as \({\text{CV}} = {\text{SD/Mean}}\); the coefficient of error (CE) was calculated as \({\text{CE}} = {{\text{CV}} \mathord{\left/ {\vphantom {{\text{CV}} {\sqrt {{\text{number }}\;{\text{of}}\;{\text{boluses}}} }}} \right. \kern-0pt} {\sqrt {{\text{number }}\;{\text{of}}\;{\text{boluses}}} }}\); precision was calculated as \({\text{precision}} = 2*{{\text{CV}} \mathord{\left/ {\vphantom {{\text{CV}} {\sqrt {{\text{number }}\;{\text{of}}\;{\text{boluses}}} }}} \right. \kern-0pt} {\sqrt {{\text{number }}\;{\text{of}}\;{\text{boluses}}} }}\); and the least significant change (LSC) was calculated as \({\text{LSC}} = {\text{CE}}*1.96*\sqrt 2\) [14]. For all analyses, a significance level of 0.05 has been used. Due to the explorative manner of this study, no adjustment for multiple comparisons has been made.

Results

Higher variability of TPTD measurements at HDP

A comparison of the single measurements within a triplicate revealed a higher variability of the measurement results and thus a lower precision for CI and GEDVI at HDP compared with HDO or HDT as evidenced by an increased SD; this was not the case for EVLWI measurements (Fig. 1).

Fig. 1
figure 1

Mean differences in SD for CI, GEDVI, EVLWI. Shown are the differences in SD (with confidence intervals) between the triplicate measurements for the parameters cardiac index (CI) (a), global end-diastolic volume index (GEDVI) (b) and extravascular lung water index (EVLWI) (c) for paused versus during (HDP/HDO) (filled triangle), during versus stopped (HDO/HDT) (filled square) and paused versus stopped HD (HDP/HDT) (filled circle)

Subsequently, the mean differences for CI and GEDVI between the single measurements were calculated, to test if the first measurement after pausing the HD should be rejected. The mean absolute differences between first and second measurements were smaller than the mean absolute differences between the second and third measurements, and therefore, the first measurement after pausing the SLED could not account for the higher variability. The mean difference for the EVLWI showed no significant difference between the settings (Table 2).

Table 2 Mean absolute differences between the first, second and third measurements for CI, EVLWI, GEDVI at HDP

TPTD measurements and catheter location

There was no significant evidence of an effect of dialysis catheter location on any of the outcome variables (Fig. 2, Table 3). The mean absolute values for HDO and HDT were in the same range, whereas the results differed from HDP. Figure 2 shows that the mean values were comparable for both locations without neither a significant nor meaningful difference. For example, the CI at HDO in the SVC (CISVC = 3.08 ± 0.21 l/min/m2) was similar to the one measured in the IVC (CIIVC = 3.19 ± 0.22 l/min/m2). Comparable results were found for the EVLWI and GEDVI and additionally in the two other settings (HDP, HDT).

Fig. 2
figure 2

Boxplots CI, EVLWI, GEDVI. Shown are the values (median/25. and 75. percentile) for the parameters cardiac index (CI) (a), extravascular lung water index (EVLWI) (b) and global end-diastolic volume index (GEDVI) (c) at ongoing HD (HDO), paused HD (HDP) and terminated HD (HDT) for measurements via a catheter in the superior vena cava (SVC) or inferior vena cava (IVC). The given p values indicate the differences between SVC and IVC

Table 3 Estimated differences and 95% confidence intervals for CI, GEDVI and EVLWI in each setting between time points and between catheter locations

CI and blood flow rate

The addition of blood flow rate to the multilevel models did not significantly alter any of the measured outcome variables (p = 0.784 for CI, p = 0.600 for GEDVI, p = 0.237 for EVLWI).

TPTD measurements for HDO, HDP and HDT

A summary of the TPTD measurements for ongoing (HDO), paused (HDP) and terminated (HDT) dialysis is given in Fig. 2. The results comparing the measurements between time points are presented in Table 3. The pausing of SLED led to significantly increased values of CI and GEDVI compared to HDO and HDT and significantly decreased values of EVLWI. Although statistical significance was also reached for the comparisons of CI and EVLWI between HDO and HDT, these differences are not considered to influence the outcome of the TPTD as they are with about 10% below or around the level of LSC (CISVC/IVC 10–15%, ELWISVC/IVC 8–21%).

Discussion

Our study shows that (1) ongoing SLED does not influence TPTD measurements, (2) pausing SLED for TPTD significantly influenced results due to a higher variability of measurements, and (3) results were not influenced by dialysis catheter site (SVC vs IVC) and blood flow rate.

SLED is frequently used instead of other better evaluated continuous RRTs, e.g., CVVH(D/F). We compared TPTD during/at paused SLED to measurements, which were taken under standard conditions after terminated SLED and after blood re-transfusion, regaining a steady state. The results of TPTD measurements during ongoing SLED were comparable to the results of the measurements after terminated SLED, while TPTD measurements assessed at paused SLED significantly deviated from ongoing and terminated SLED.

Although we have detected statistically significant differences in CI and EVLWI between ongoing and terminated SLED in combination with a nonsignificant difference in GEDVI, the clinical impact of these differences is negligible. The precision of the triplicate measurements at ongoing, paused and terminated SLED for CI is ≤ 10% and thus acceptable. For GEDVI and ELWI, some values at paused (GEDVI), respectively, terminated SLED (ELWI) are above this limit and thus less precise. The differences in, e.g., CI between ongoing/terminated and paused SLED are statistically significant, but still under the threshold of 15% [15,16,17] to be clinically relevant as well as below the threshold of LSC [14]. For differences below the threshold of LSC, it is not sure that the changes are true or linked to the error of the technique [14]. In contrast considering the absolute difference in CI of 0.4 L/min/m2 for ongoing versus terminated SLED in clinical practice, this would unlikely cause a change in the existing treatment regime, as opposed to a difference in CI of 2.0 L/min/m2 (paused vs ongoing SLED) or 1.7 L/min/m2 (paused vs terminated SLED), respectively.

In contrast, pausing SLED for the duration of TPTD measurements significantly influences TPTD measurements presumably by inducing turbulences or temperature differences and fluctuations, but neither dialysis catheter position nor blood flow rate had significant influence on TPTD variables. TPTD is thus considered sufficiently accurate during ongoing SLED as well as after its termination although one would expect higher temperature fluctuations than with continuous RRTs. Our results show that TPTD measurements differ significantly depending upon whether they are obtained during or shortly after terminating SLED compared with pausing SLED, indicating that pausing SLED to perform TPTD measurements results in inaccurate and imprecise estimates of cardiac index and other important hemodynamic variables.

Investigators have found variable effects of continuous renal replacement therapy on TPTD measurements, with different findings depending upon RRT catheter location, flow rates, and whether RRT is ongoing, paused or stopped during the TPTD measurements. Our findings suggest that pausing RRT results in imprecise and inaccurate TPTD-derived hemodynamic measurements, contributing to the existing body of knowledge about the effects of RRT on TPTD hemodynamic measurements. Our study expands existing knowledge because we focus on SLED, whereas previous studies have examined the effects of CVVH(D) on TPTD measurements with controversial results (Table 4).

Table 4 Studies investigating the influence of running RRT and/or catheter location on TPTD

At first, some authors doubted whether TPTD at ongoing RRT delivers valid data [4,5,6,7,8]. They detected significant changes in the CO [5]/CI [4, 6,7,8], ITBVI [4, 5]/GEDVI [6, 8] and EVLWI [5, 6, 8] at ongoing RRT depending on the dialysis catheter location (catheter behind temperature detector tip versus between bolus injection and temperature detector) and therefore did not advise to measure during ongoing RRT [4,5,6, 8] or at least recommend a careful interpretation of these data [7].

In contrast, other authors [10,11,12] systematically evaluated the optimal time point of TPTD measurements and detected significant changes between measurements at ongoing versus paused/disconnected/no RRT. In line with our results, these authors concluded that small changes in measurements are clinically negligible and deemed TPTD measurements at ongoing RRT to be valid not justifying pausing the dialysis system during measurement. In one of these studies [12], a quintuplicate measurement was applied, and the authors recommended: “If system should be paused, at least reject the first measurement or wait until blood temperature has normalized.” In our study, we could also show that the variability for the CI and GEDVI of the triplicate measurements at paused RRT was higher than ongoing RRT. We calculated the mean differences between the first and the two other measurements, but did not find a meaningful difference between the first and consecutive measurements. Nevertheless, we also detected that the change from running to paused RRT influenced the variability of TPTD results, and therefore, time should be given allowing to reach a steady state before TPTD measurements after pausing/stopping the RRT.

In all these studies, a continuous RRT (CVVH(D)) was applied, whereas in the present study, the effect of SLED on TPTD was evaluated. Due to the higher transfer volume of cold substitute (9 L/h of dialysate in SLED versus 2 L/h in CVVH) as compared to other RRT procedures, one would expect more pronounced temperature changes that might interfere with TPTD measurements. In addition, there is no active heating in the SLED, which leads to the cooling of the dialysate by 0.5 °C/h, implying that further temperature fluctuations are difficult to calculate. We also investigated the optimal time to recalibrate the system while SLED is already in progress. To the best of our knowledge, only two studies also systematically addressed the reliability of TPTD during SLED. Pathil et al. [7] measured every 8 h during phases of hemodynamic stability without hemodialysis and immediately after onset of SLED. They detected a significant, but clinically acceptable decrease in CI, a significant decrease in GEDVI and a discrete reduction in EVLWI and concluded that measurements during SLED should be carefully interpreted when relying solely on these. They compared measurements before and immediately after onset of the SLED, but this is not comparable to our approach and does not answer if recalibration of TPTD should be performed at ongoing or paused SLED. Huber et al. [3] performed TPTD immediately before and 5 min after connection to SLED, as well as immediately before and after disconnection of SLED and re-transfusion, whereas we measured while ongoing, at interrupted and after terminated SLED and re-transfusion of blood (Fig. 3). Comparable to our findings, TPTD results in these studies seem relatively independent of the presence of a running RRT in a steady state, despite the presumed higher temperature fluctuation using SLED. Accordingly, Huber et al. concluded that TPTD is feasible at ongoing RRT, but, in contrast to our study, they did not separately evaluate if pausing has an influence on TPTD. This aspect had already been studied by Sakka, Dufour and Heise [3, 10,11,12], who found no [11] or a clinically irrelevant [10, 12] difference between running RRT and paused RRT, although data are only available considering CVVH(D) techniques. Our data add some further knowledge also considering effects of pausing SLED, as an increasingly used RRT technique, on TPTD measurements.

Fig. 3
figure 3

Comparison of the measurement time points to investigate the influence of ongoing SLED on TPTD results between Huber et al. 2016 and our study. Huber et al. 2016 performed their TPTD measurements before connection of the pump (T1), during SLED with pump on (T2), during SLED with pump on after pausing the RRT (T3) and after disconnection (T4). We performed our measurements during SLED (HDO), at paused SLED (pump off/on, HDP) and after termination (HDT)

Huber’s and our study show that the changes in CI found during RRT did not influence the TPTD results in a clinically relevant manner. However, in theory the extracorporeal blood flow might lead to a loss of the injected bolus and thus may falsify the determination of the pumped blood volume of the heart by time, defined as CO or CI [per body surface area (BSA)]. The higher the blood flow, the higher (in theory) the loss of the bolus volume, thus leading to an overestimation of the CO up to a maximum of 300%. A ratio (blood flow to CO) below 0.5 seems not to affect TPTD accuracy [18]. In our study, there was a detectable, but clinically not meaningful correlation between the blood flow rate and the CI, considering different settings (HDO, HDP, HDT). Additionally, we calculated the ratio between the blood flow and the CO and compared it with the study of Sakka, Dufour and Heise, using CVVHF (Table 5). The calculated ratios were between 0.01 and 0.06, far below 0.5, and therefore indicating that the accuracy of TPTD was not affected by different blood flow rates. The blood flow rate in our patients was relatively high in combination with a comparably low CI, and though the blood flow/CO ratio was relatively high, we did not detect a negative impact on TPTD.

Table 5 Calculation of the ratio between the blood flow and CO

The location of the dialysis catheter in relation to the CVC, via which the cold bolus is applied, might also influence the TPTD measurements [10, 14,15,16]. In the situation when TPTD is performed in combination with RRT, the proximity of the dialysis catheter to the CVC suggests that this might falsify TPTD results by immediate extraction of the cold bolus. Sakka et al. 2007 described no evidence that this effect relevantly influenced their results. They measured at low blood flow rates, and they did not see any influence of the dialysis catheter position either whether it was placed in the same vessel as the CVC or not [10]. Dufour et al. reported that the paused RRT significantly increased the blood temperature but did not influence the temperature difference between the blood in the SVC and the femoral artery. The authors postulated that the low extracorporeal blood flow rate did not induce a significant alteration in the thermal effect of the cold boluses [11]. Martinez-Simon et al. detected a two-peak temperature curve in the femoral artery during RRT after injection of the cold bolus. They inserted the bolus into the third lumen of a dialysis catheter and did not use a separate central line for PiCCO™. They concluded that the bolus might have directly been aspirated by the hemofiltration flank and therefore been re-injected later [4].

In our study, the cold bolus was injected into a separate CVC avoiding a possible aspiration and we could demonstrate that the location of the dialysis catheter did not relevantly influence the TPTD measurements. Although the number of patients with the catheter placed in the SVC is more than twice compared to the number of patients with the catheter placed in the IVC in our study, the results are statistically tested for equivalence and reliability. These results are in line with those of the studies of Sakka et al. and Dufour et al; both did not detect a difference in TPTD measurements when the dialysis catheter tip is placed in the SVC as compared to IVC [10, 11]. Van Craenenbroeck et al. [8] described similar results, but observed a more pronounced difference between TPTD results with and without RRT when the PiCCO™ catheter was placed in the IVC.

There are some important limitations of our study. First, we did not validate our TPTD measurements against another method to set this as a gold standard. In addition, we did not validate TPTD-derived CI with another method for the measurement of cardiac output, which would not be affected by SLED. Using echocardiography of aortic blood flow as an example would not likely be affected by the temperature or flow rate of blood returning from the SLED device into the vena cava or the right atrium. Hence, we are unable to conclude if a systematic bias introduced by SLED favors either paused, ongoing or terminated SLED during TPTD measurement. Second, we did not measure the body and blood temperature. A limitation of TPTD measuring during RRT could be a decreased accuracy of the measurements as they are highly relying on the blood volume and temperature. Even small changes in the indicator volume might falsify the results of the hemodynamic monitoring. This liability of measurement accuracy on minimal variations in the blood temperature implies to test variations in the indicator bolus such as cooled versus room temperature saline or the replacement of the saline with a lithium dilution. Since neither the body nor the blood temperature was evaluated before bolus injection or at the time of pausing and restarting the dialysis, these effects on TPTD measurements could not be tested. It would be interesting to evaluate if a longer waiting period before measurement may improve the accuracy of measurements compared to the presumably true value obtained after stopping the dialysis. Furthermore, our conclusions might only apply to patients with the CVC for TPTD measurements placed in the SVC. Results could be different when the catheter is inserted in the femoral vein, especially in cases with abdominal hypertension. Finally, the data are of observational nature with a small number of patients, in particular concerning the small number of patients with location of the dialysis catheter in the IVC, which resulted from the fact that the femoral access route is not common in our clinic, and independently thereof repeated measurements within many patients. As this study tried to mirror every day clinical practice and as it is an exploratory study, we aimed to investigate if larger (clinically relevant) differences between the TPTD measurements are to be expected for ongoing versus interrupted SLED. From this, we did no sample size calculation a priori. Since there was a clear statistical difference between measurement conditions (even in this small sample size), a sample size calculation is not necessary for this exploratory study.

Conclusions

TPTD measurements are not clinically meaningful influenced by ongoing SLED, regardless of blood flow and location of the dialysis catheter. However, starting or stopping the dialysis temporarily affected TPTD results. In a real-life scenario, SLED is running for up to 12 h and a TPTD measurement/recalibration of the PiCCO™-system needs to be performed regularly or in case of hemodynamic deterioration during this process. The easiest and most efficient way to perform these measurements in clinical routine would be during ongoing SLED, or, if this would deliver falsified results, at paused SLED. We have shown that measurements at ongoing SLED deliver accurate results, and as this is the most convenient way to perform routine measurements, our data therefore suggest that the dialysis system needs not to be stopped for the TPTD measurements.

Abbreviations

AKI:

acute kidney injury

BSA:

body surface area

CE:

coefficient of error

CI:

cardiac index

CV:

coefficient of variation

CVC:

central venous catheter

CVVH(D/F):

continuous veno-venous hemofiltration/hemodialysis/hemodiafiltration

EVLWI:

extravascular lung water index

GEDVI:

global end-diastolic volume index

HD:

hemodialysis

HDO:

ongoing hemodialysis

HDP:

paused hemodialysis

HDT:

terminated hemodialysis

IVC:

inferior vena cava

LSC:

least significant change

MAP:

mean arterial pressure

MOF:

multiple organ failure

RRT:

renal replacement therapy

SD:

standard deviation

SLED:

sustained low efficiency dialysis

SVC:

superior vena cava

SVRI:

systemic vascular resistance index

TPTD:

transpulmonary thermodilution

References

  1. Reuter DA, Huang C, Edrich T, Shernan SK, Eltzschig HK. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010;110(3):799–811.

    Article  Google Scholar 

  2. Marx G, Schuerholz T, Sumpelmann R, Simon T, Leuwer M. Comparison of cardiac output measurements by arterial trans-cardiopulmonary and pulmonary arterial thermodilution with direct Fick in septic shock. Eur J Anaesthesiol. 2005;22(2):129–34.

    Article  CAS  Google Scholar 

  3. Huber W, Fuchs S, Minning A, Kuchle C, Braun M, Beitz A, et al. Transpulmonary thermodilution (TPTD) before, during and after sustained low efficiency dialysis (SLED). A prospective study on feasibility of TPTD and prediction of successful fluid removal. PLoS One. 2016;11(4):e0153430.

    Article  Google Scholar 

  4. Martinez-Simon A, Monedero P, Cacho-Asenjo E. Erroneous measurement of haemodynamic parameters by PiCCO monitor in a critically ill patient with renal replacement therapy: a case report. Crit Care. 2006;10(2):410.

    Article  Google Scholar 

  5. Mason N, Froude A, Holst B, Saayman A, Findlay G. The effect of continuous veno-venous haemofiltration on PiCCO(®)haemodynamic parameters. Critical Care. 2005;9(Suppl 1):P65.

    Article  Google Scholar 

  6. Neirynck V, Willems A, Peeters D, Van Regenmortel N, De laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration on transpulmonary thermodilution-derived parameters. Critical Care. 2009;13(1):P274.

    Article  Google Scholar 

  7. Pathil A, Stremmel W, Schwenger V, Eisenbach C. The influence of haemodialysis on haemodynamic measurements using transpulmonary thermodilution in patients with septic shock: an observational study. Eur J Anaesthesiol. 2013;30(1):16–20.

    Article  Google Scholar 

  8. van Craenenbroeck A, VanIngelgem A, Palmers PJ, Van Regenmortel N, De Laet I, Schoonheydt K, et al. Influence of continuous venovenous hemofiltration (CVVH) and catheter position on transpulmonary thermodilution derived parameters with PiCCO. Intensive Care Med. 2010;36(Suppl 2):S294.

    Google Scholar 

  9. Huber W, Rockmann F. Invasive und nichtinvasive Möglichkeiten des haemodynamischen Monitorings. Intensivmed. 2008;45:337–59.

    Article  Google Scholar 

  10. Sakka SG, Hanusch T, Thuemer O, Wegscheider K. The influence of venovenous renal replacement therapy on measurements by the transpulmonary thermodilution technique. Anesth Analg. 2007;105(4):1079–82.

    Article  Google Scholar 

  11. Dufour N, Delville M, Teboul JL, Camous L, Favier du Noyer A, Richard C, et al. Transpulmonary thermodilution measurements are not affected by continuous veno-venous hemofiltration at high blood pump flow. Intensive Care Med. 2012;38(7):1162–8.

    Article  Google Scholar 

  12. Heise D, Faulstich M, Morer O, Brauer A, Quintel M. Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques. Minerva Anestesiol. 2012;78(3):315–21.

    CAS  PubMed  Google Scholar 

  13. R Core Team. R: A language and environment for statistical computing. R Foundation for Statistical Computing. Vienna, Austria 2014.

  14. Monnet X, Persichini R, Ktari M, Jozwiak M, Richard C, Teboul J-L. Precision of the transpulmonary thermodilution measurements. Crit Care. 2011;15(4):R204.

    Article  Google Scholar 

  15. Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL. Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010;14(3):R109.

    Article  Google Scholar 

  16. Monnet X, Osman D, Ridel C, Lamia B, Richard C, Teboul JL. Predicting volume responsiveness by using the end-expiratory occlusion in mechanically ventilated intensive care unit patients. Crit Care Med. 2009;37(3):951–6.

    Article  Google Scholar 

  17. Jabot J, Teboul JL, Richard C, Monnet X. Passive leg raising for predicting fluid responsiveness: importance of the postural change. Intensive Care Med. 2009;35(1):85–90.

    Article  Google Scholar 

  18. Haller M, Zollner C, Manert W, Briegel J, Kilger E, Polasek J, et al. Thermodilution cardiac output may be incorrect in patients on venovenous extracorporeal lung assist. Am J Respir Crit Care Med. 1995;152(6 Pt 1):1812–7.

    Article  CAS  Google Scholar 

Download references

Author’s contributions

SG treated the patients, collected the data and wrote the manuscript. LS performed the statistical analysis. CR made substantial contributions to study design and interpretation of the data and reviewed the manuscript. SS treated the patients and reviewed the manuscript. FE made substantial contributions to study design and reviewed the manuscript. All authors read and approved the final manuscript.

Acknowledgements

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Funding

The authors received no funding for this work.

Availability of data and materials

The dataset supporting the conclusions in this article is owned by the participating study centers. Upon request, researchers may get access to the data at the study centers at the Department of Clinical Toxicology, Klinikum rechts der Isar, Technical University of Munich.

Consent for publication

No details, images or videos relating to individual patients are shown; therefore, no consenting beyond study participation is needed.

Ethics approval and consent to participate

The study protocol was approved by the ethics committee of the Faculty for Medicine of the Technical University of Munich (ethics no. 159/14).

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Stefanie Geith.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Geith, S., Stecher, L., Rabe, C. et al. Sustained low efficiency dialysis should not be interrupted for performing transpulmonary thermodilution measurements. Ann. Intensive Care 8, 113 (2018). https://doi.org/10.1186/s13613-018-0455-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13613-018-0455-x

Keywords